EasyCare MANGAR AIRFLO 24 Скачать руководство пользователя страница 10

ENGLISH

AIRFLO 24

10

Maintenance

 - fig.  

4

The  Airflo  24  has  been  designed  to  be  maintenance  free  and 

apart from replacing the battery, the Airflo contains no other user 

serviceable parts.

Battery replacement:

Disposal

When the Airflo

 

24 has reached the end of it’s working life, remove the battery for safe disposal in 

accordance with local regulations, and dispose of the Airflo

 

24 at your local household waste recycling 

centre (civic amenity site).

Cleaning

Unless contamination is suspected, periodic cleaning is all that is recommended. The Airflo

 

24 should be 

cleaned/disinfected as follows:

Unplug the Airflo

 

24 from the battery charger. Clean the external surfaces using a non abrasive damp

cloth with warm water. Wipe dry thoroughly with a dry towel before reconnecting to the battery

charger. Do not allow moisture to enter any part of the compressor.

Do not use solvent-based products.

Disinfecting/Sterilising

Disinfection – Pre-clean the Airflo compressor by wiping with a damp cloth and then disinfect by

wiping with a 1:10 solution of household liquid bleach and water - 1 part bleach to 10 parts cold

water (10,000ppm (1%) Sodium Hypochlorite solution). Do not allow moisture to enter any part of

the compressor. Allow to air dry. In usual circumstances this product does not require sterilisation for

reuse.

Storage

When not in use, store in a dry place together with these instructions. To maintain the batteries, when

being stored for long periods, leave the Airflo 24 on charge.

When it becomes apparent that the Airflo

 

24 is producing fewer than expected lifting cycles, the battery is 

probably reaching the end of its working life and should be replaced. For optimal performance, Winncare 

recommends replacing batteries every 12 months.

If a fault should develop or if you require assistance in setting up using or maintaining the equipment, please 

contact your supplier.

A

B

4

Turn the Airflo 24 onto its end containing the air outlet socket,

remove the screw (

A

) and slide the end panel (

B

) away from 

the housing.

Disconnect the wire from the battery pack and remove the

battery pack from the battery compartment.

Fit the replacement battery pack, ensuring that the connector

is secure, refit the end panel and secure the screw (

A

).

To calibrate the battery indicator, it is essential before use to

charge for 12 hours.

www.easycaresystems.co.uk

Содержание MANGAR AIRFLO 24

Страница 1: ...USER INSTRUCTIONS MANGAR AIRFLO 24 COMPRESSOR USER INSTRUCTIONS www easycaresystems co uk...

Страница 2: ...HOURS before first use BEFORE USE 1 x Airflo 24 compressor A fitted with a 24v 2200mAh Ni MH battery pack 1 x Battery charger with mains adaptors B 1 x User Instructions 1 x Mains Power Adaptor Option...

Страница 3: ...the case of a power cut switching the power socket off or disconnecting the charging lead the charging process will not be interrupted as the unit remembers the previous charge state OPERATING CHARGE...

Страница 4: ...in the battery Battery Status Indicator Button UNDERSTANDING THE BATTERY STATUS INDICATOR LIGHTS TO DETERMINE THE BATTERY CHARGE STATE YOU MUST DISCONNECT THE CHARGING LEAD OR REMOVE THE AIRFLO 24 FR...

Страница 5: ...battery state The hand control button is then released when a compartment in the lifting device is filled Product will then automatically turn off after 2 5 mins of inactivity 4 GREEN OK to Use GREEN...

Страница 6: ...use the Airflo 24 and its connected hoses are laid out in such a way as to avoid any risk of tripping or strangulation Ensure that children and pets do not interfere with the safe operation of the Air...

Страница 7: ...ge the battery When the compressor is running the battery status indicator lights will only indicate the state of battery charge as it was at the start of the lift When the compressor stops running th...

Страница 8: ...will flash green After 3 hours the first light will become a steady green and the second light will start flashing After 6 hours the bottom 2 lights will remain green and the third light will flash Af...

Страница 9: ...irectly from a mains power socket The adaptor is specific to the Airflo 24 and is the only approved unit to be used for this purpose Connect the adaptor between a mains power socket and the adaptor in...

Страница 10: ...water 1 part bleach to 10 parts cold water 10 000ppm 1 Sodium Hypochlorite solution Do not allow moisture to enter any part of the compressor Allow to air dry In usual circumstances this product does...

Страница 11: ...ck that there are no kinks in the air hose Compressor keeps stopping when trying to inflate the lifting aid Check that there are no kinks in the air hose No Battery status indicator Lights Unit switch...

Страница 12: ...ime O Non operating time Max continuous operating time 5 minutes For indoor use only Do Not connect to mains power in a bathroom or outdoors Recycle Do not dispose of in household waste Type BF equipm...

Страница 13: ...he following statements do not affect the consumer s statutory rights This product is sold on the understanding that in the event of any defect of manufacture or material appearing within the warranty...

Страница 14: ...Special Products and Service Contact Us Units 5 and 6 Lower Vicarage Road Southampton Hampshire SO19 7RJ T 02380 420678 E info easycaresystems co uk...

Отзывы: