12.

 

Trouble shooting 

 

 
Problem 

Possible cause 

Corrective action 

Piston with difficulty to 
move 

Dirt, possible formation of crystals 

Stop dispensing immediately. 
Loosen the piston by circular 
motion without disassembling. 
Follow all the cleaning 
instructions.  

Filling not possible 

-Volume adjusted to minimum setting 
-Sticking filling valve 

-Set the required volume 
-Clean the filling valve and 
replace with sealing washer if 
necessary. 

Instrument does not 
dispense liquid 

- Closing cap is on  
-Discharge tube not firmly connected 
or damaged 

-Remove closing cap  
-Push on the closing cap. Hold 
the piston in place and push 
the original 

manufacturer’

discharge tube on firmly.  
Replace a deformed or 
damaged discharge tube.  

Air bubbles in the 
instrument 

- Reagent with high vapor pressure 
has been poured too quickly. 
 
-Seal not inserted, knurled locking 
nuts not firmly connected 
 
 
 
 
 
-Air not expelled from the instrument 
-Filling tube not firmly connected or 
damaged     
 
 
-Valves dirty, not firmly installed, or 
damaged                                     
 

-Pour the reagent slowly. 
 
 
-Double check that seal (12) is 
in place, and that the knurled 
locking nuts at the discharge 
tube are firmly seated and 
finger tightened 
-Priming the instrument 
 
-Push the filling tube on firmly. 
If necessary, cut off 
approx. 1 cm of tube at the 
upper end and reconnect it 
or replace filling tube. 
-Cleaning procedure. Tighten 
the valves using the mounting 
tool. If necessary, replace the 
valves and 
sealing washers. 

Dispensed volume is too 
low 

-Discharge tube not firmly connected  
or damaged 
-Filling tube not firmly connected or 
damaged 
 
 
 
 
-Valves not firmly installed or 
damaged 
 
 
-Sticky filling valve 
 

-Push the original 

manufacturer’s tube on firmly.

 

-Cleaning procedure. Push the 
filling tube on firmly. If this 
does not suffice, cut off 
approx. 1 cm of tube at the 
upper end and re-connect it or 
replace filling tube. 
-Clean the instrument. Tighten 
the valves using the mounting 
tool, if necessary, replace the 
valves and sealing washers.  
- Clean or replace discharge 
tube, if necessary. 

Содержание EASY 20K+

Страница 1: ...Bottle top dispenser EASY 20K Please read the User Manual carefully before use and follow all operating and safety instructions user manual english espa ol...

Страница 2: ...tomer service representative with the following information Serial number on the back side Description of the problem Your contact information Warranty This instrument is guaranteed to be free from de...

Страница 3: ...hes and protect yourself adequately Do not operate this instrument in any manner not described in this User manual In case of abnormal circumstances during the manipulation of this device please stop...

Страница 4: ...llowing limits 15 to 40 C of instrument and reagent Vapor pressure up to 1000 mbar Density up to 2 2 g cm3 4 Materials When the instrument is correctly used the dispensed liquid comes into contact wit...

Страница 5: ...cult to move never use force Immediately stop dispensing and follow cleaning instructions or contact the manufacturer Note Check the compatibility table or contact the manufacturer for special applica...

Страница 6: ...tube opening Hold an appropriate collecting vessel under the discharge tube opening Gently pull the piston up then press it down rapidly Repeat this procedure until most of the air bubbles in the gla...

Страница 7: ...arge tube and closure cap 8 1 Setting the volume Rotate the volume adjustment knob counterclockwise to unlock it Move the knob along the scale to target volume Rotate the knob clockwise to lock it at...

Страница 8: ...Pull and push the piston for several time to dispense the remaining reagent from the discharge tube into the reagent bottle as much as possible 5 Remove the dispenser from the empty bottle to another...

Страница 9: ...the closed autoclave to avoid deformation through too rapid cooling After every autoclaving inspect all parts for deformities or damage If necessary replace them Do not reassemble the instrument unti...

Страница 10: ...instrument Filling tube not firmly connected or damaged Valves dirty not firmly installed or damaged Pour the reagent slowly Double check that seal 12 is in place and that the knurled locking nuts at...

Страница 11: ...i n al Cliente la siguiente informaci n N mero de serie del equipo en el panel trasero Descripci n del problema detectado Sus datos de contacto Garant a Este equipo est garantizado contra cualquier de...

Страница 12: ...jase adecuadamente No opere este instrumento de ninguna manera no descrita en este manual de usuario En caso de circunstancias anormales durante la manipulaci n de este dispositivo por favor detenga e...

Страница 13: ...C de instrumento y reactivo Presi n de vapor hasta 1000 mbar Densidad hasta 2 2 g cm3 4 Materiales Cuando el instrumento se utiliza correctamente el l quido dispensado entra en contacto nicamente con...

Страница 14: ...e la fuerza Deje de dispensar inmediatamente y siga las instrucciones de limpieza o comun quese con el fabricante Nota Consulte la tabla de compatibilidad o comun quese con el fabricante para aplicaci...

Страница 15: ...ipiente colector adecuado debajo de la abertura del tubo de descarga Tire suavemente del pist n hacia arriba y luego presi nelo hacia abajo r pidamente Repita este procedimiento hasta que se hayan eli...

Страница 16: ...la perilla de ajuste de volumen en sentido contrario a las agujas del reloj para desbloquearla Mueva la perilla a lo largo de la escala hasta el volumen objetivo Gire la perilla en el sentido de las...

Страница 17: ...ivo a otra botella vac a Tire y empuje el pist n varias veces para dispensar el reactivo restante del tubo de descarga a la botella de reactivo tanto como sea posible 5 Retire el dispensador de la bot...

Страница 18: ...se deforme por un enfriamiento demasiado r pido Despu s de cada esterilizaci n en autoclave inspeccione todas las piezas en busca de deformidades o da os Si es necesario reempl celos No vuelva a monta...

Страница 19: ...nectado firmemente o da ado V lvulas sucias mal instaladas o da adas Vierta el reactivo lentamente Verifique dos veces que la junta est en su lugar y que las tuercas moleteadas de bloqueo en el tubo d...

Страница 20: ...e en vertederos De conformidad con la Directiva 2002 96 CE los usuarios de la Uni n Europea de aparatos el ctricos y electr nicos tienen la oportunidad de retornar el instrumento para su eliminaci n a...

Страница 21: ...www labbox com...

Отзывы: