5
OPERATING INSTRUCTIONS
CAUTION
Inspect the Easy Vac Aspirator for visual damage, DO NOT USE
if damaged.
1. Place Easy Vac Aspirator near electrical receptacle on a table,
floor or other horizontal, smooth, hard surface.
Allow a minimum of 4 in. (10 cm) clearance on all sides to ensure
adequate air flow for cooling of Aspirator.
2. Press down on Aspirator until Suction Feet are secure.
3. Be sure the On/Off Switch is in the “OFF”
position.
4. Be sure the Inline Hydrophobic Filter is properly connected to the
vacuum port on the Suction Canister lid.
• The Inline Hydrophobic Filter prevents liquid from entering
the vacuum pump.
CAUTION
dO NOT
run the Easy Vac Aspirator without Inline Hydrophobic
Filter. This will void the warranty.
5. Plug the Aspirator into receptacle.
6. Turn the On/Off Switch to the “
ON
” position.
7. Block the patient connection on the Suction Canister and adjust
the Vacuum Control Knob to the desired vacuum.
To INCREASE vacuum - Turn knob CLOCKWISE
To DECREASE vacuum - Turn knob COUNTERCLOCKWISE
• Vacuum should be set as prescribed by Provider.