Precision Medical, Inc.

300 Held Drive

Northampton, PA 18067 USA

Toll Free Tel: 800-272-7285

Toll Free Fax: 800-353-1240

Tel:(+001) 610-262-6090

Fax:(+001) 610-262-6080

1907

 Rev10  08/30/12 (?M) 

P

rinted

 

in

 USA

Содержание PM60

Страница 1: ...S CAUTION Federal USA law restricts this device to sale by or on the order of a physician 300 Held Drive Tel 001 610 262 6090 Northampton PA 18067 USA Fax 001 610 262 6080 ISO 13485 Certified www prec...

Страница 2: ...f you do not understand this manual DO NOT USE the Easy Vac Aspirator and contact your Provider Explanation of abbreviations inHg Inches of Mercury mmHg Millimeters of Mercury SAFETY Information Warni...

Страница 3: ...is not suitable for use in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide WARNING To reduce risk of burns electrocution fire or injury to persons The Easy Vac Aspirat...

Страница 4: ...F 10 C to 40 C Storage Temperature Range 4 F to 104 F 20 C to 40 C Max 95 Noncondensing Filter Inline Hydrophobic Safety Systems Automatic Resetting Thermal Limit Switch Noise Level 69db 1 meter Equip...

Страница 5: ...any label becomes illegible or lost contact Precision Medical Inc Vacuum Gauge Suction Feet Control Knob On Off Switch Patient Tubing Connection Standard 800cc Suction Canister with Float Shutoff Inli...

Страница 6: ...sition 4 Be sure the Inline Hydrophobic Filter is properly connected to the vacuum port on the Suction Canister lid The Inline Hydrophobic Filter prevents liquid from entering the vacuum pump CAUTION...

Страница 7: ...nstructed by Provider 10 When procedure is completed press the On Off Switch to the off position DO NOT overfill the canister 11 Empty the Suction Canister as follows Remove the canister from its brac...

Страница 8: ...use only DO NOT clean Cleaning for Exterior Components 1 Disconnect the Aspirator from power source and other connections before cleaning 2 Use a cloth or sponge dampened with soap and water 3 Wipe d...

Страница 9: ...RNING If Aspirator is used for treatment of infectious disease consult your Physician or Healthcare Professional for recommended procedures for proper disposal Replacement Parts Description Part Patie...

Страница 10: ...acuum on gauge NO vacuum at Patient Connection Tubing 1 Hydrophobic Filter not connected properly 2 Suction Canister not connected or assembled properly 3 Defective Suction Canister 4 Defective Regula...

Страница 11: ...es about the merchandise described in this contract and any such statements shall not be relied upon and are not part of the contract for sale Thus this writing is a final complete and exclusive state...

Страница 12: ...Precision Medical Inc 300 Held Drive Northampton PA 18067 USA Toll Free Tel 800 272 7285 Toll Free Fax 800 353 1240 Tel 001 610 262 6090 Fax 001 610 262 6080 1907 Rev10 08 30 12 M Printed in USA...