ADJUSTMENTS AND REPLACEMENTS
Additional Service Procedures
06DEC05
AR4366-1
Page
211 of 219
Doing a Calibration for the Scan
Important
For systems with software version
≥
4.1,
you must have a “Session ID” for access to “Service
Functions” and “Diagnostics”. See SERVICE BULLETIN 843.
1
At the main menu, click:
•
[Service Functions]
•
[Diagnostics]
•
[Scan Calibration]
2
Enter the number from the display on the DOSIMETER. See
3
Insert the first CASSETTE into the system.
4
Click [Start Calibration Scan].
5
Insert the exposed CASSETTES every 2 minutes.
Important
If the calibration is not successful, an error code occurs.
6
Was the calibration successful?
7
At the “Last Calibrated Date” screen, locate the current date and time for the
corresponding CASSETTE.
8
Locate and correct the corresponding error code in the error log.
Note
For error codes 16019 and 16020, you must see the adjustment for the
Yes
No
Calibration is complete. Continue with