ADJUSTMENTS AND REPLACEMENTS
Additional Service Procedures
06DEC05
AR4366-1
Page
205 of 219
6
Use tape to fasten the COPPER and
ALUMINUM FILTERS TL-5259 to each
other.
Important
The COPPER FILTER must be directly next
to the X-RAY TUBE.
7
Insert both of the FILTERS into the
bottom of the X-RAY TUBE. If
necessary, use tape to adhere the
FILTERS to the bottom of the X-RAY
TUBE.
Preparing the Equipment
Important
•
A LEAD SHEET and the help of a RT is
necessary for this procedure.
•
The specification for the “mAs” might
have a difference.
1
Use the following specifications for the
first setup of the X-ray equipment:
•
80 “kVp”
±
1
•
80 “mAs”
2
Remove the CASSETTE from the floor.
H149_0013AC
X-RAY TUBE
H149_0013ACC
ALUMINUM FILTER
COPPER FILTER
tape
H149_0014AC
LEAD SHEET
H149_0014ACA
DOSIMETER