TABLE OF CONTENTS 

 

I.     INTRODUCTION ..............................  2 
II.    PANEL DESCRIPTION ....................  3 
III.   OPERATING INSTRUCTIONS ........  4 
A.   MEASUREMENT PROCEDURES ...  4 

1.   Air Velocity Measurement .............  4 
2.   Thermocouple Measurement ........  4 
3.   RH & Temperature Measurement .  5 
4.   Light  Measurement  .......................   5 

B.    ADDITIONAL FUNCTIONS ..............  5 

1. Data Hold ......................................  5 
2. Max/Min ........................................  6 
3. Automatic Shut Off ........................  6 
4. Overload Indicator .........................  6 
5.  Battery Replacement ....................  6 

IV.   SPECIFICATIONS ...........................  7 
V.   PRECAUTIONS ...............................  8 
VI.   WARRANTY .....................................  8 
 

I.  INTRODUCTION 

For environmental testing anywhere.  
Combines a light meter, humidity 
meter, thermometer & anemometer 
into a single compact unit.  Features 
touch-tone buttons, max-min & hold 
functions.  Comes ready to use with 
wrist strap, instructions, 9V battery, 
& soft carrying case. 

Eastern Energy Co., Ltd.

 

40/4 Vitoondumri Rd., Banbueng

   

Banbueng, Chonburi, Thailand

 

 

Tel: 66-3844-6117

[email protected]

www.eastern-energy.com

Содержание 850070

Страница 1: ...Mini Environmental Quality Meter 850070 Instruction Manual Eastern Energy Co Ltd 40 4 Vitoondumri Rd Banbueng Banbueng Chonburi Thailand Tel 66 3844 6117 sale ete co th www eastern energy com...

Страница 2: ...6 4 Overload Indicator 6 5 Battery Replacement 6 IV SPECIFICATIONS 7 V PRECAUTIONS 8 VI WARRANTY 8 I INTRODUCTION For environmental testing anywhere Combines a light meter humidity meter thermometer a...

Страница 3: ...T CD button 6 FUNCTION button 7 Air Flow Sensor 8 Thermocouple Input Socket 9 RH Sensor 10 Light Sensor 11 LCD Display 12 Wrist Strap 7 9 11 2 3 5 6 10 12 8 1 4 Fig 1 Eastern Energy Co Ltd 40 4 Vitoon...

Страница 4: ...re and the unit of measure are displayed on the LCD 2 Thermocouple Measurement Plug a Type K thermocouple probe not included in the THERMOCOUPLE INPUT SOCKET making sure that the that the polarity is...

Страница 5: ...it of measure Point the LIGHT SENSOR fig 1 10 toward the light source to be measured The results are shown on the LCD The overload indicator is displayed if the range is exceeded The zero point may dr...

Страница 6: ...reverts to the current reading 3 Automatic Shut Off The instrument has an automatic shut off function in order to prolong battery life After approximately 10 minutes without activity no buttons pushe...

Страница 7: ...Range Res Accuracy C ambient 0 50 0 1 1 2 C F ambient 32 122 2 5 F C Type K 100 1300 1 1 C F Type K 148 2372 1 2 F RH 10 95 6 30 60RH otherwise 8 Airspeed m s 0 4 25 0 3 2d Airspeed km h 1 4 90 0 Airs...