3

Note:

What showed in the above figure is front panel terminal, and rear panel terminal is

basically the same as the front panel terminal but without 12A-current shift, with
current-protection fuse on the rear panel.
Protection limits of input terminal are as follows:

Main input (HI and LO) terminals.

HI and LO input terminals are used in voltage, resistance, capacitance, continuity, frequency and
diode test, and the protection limits of the two terminals are as follows:
1) Protection limit from HI to LO. Protection limit from HI to LO is 1000 VDC or 750 VAC,
which is the maximum measurable voltage. The limit can also be expressed as 1000Vpk at most.
2) Protection limit from LO to grounding. Relative to grounding LO input terminal can safely
“float” to 500Vpk at most.

Protection limit of sense terminals

HI and LO sense terminals are only used in 4-wire resistance and ratio measurement. Protection
limit for all the terminal pairs is 200Vpk, and all the terminal pairs refer to LO SENSE to LO
INPUT, HI SENCE to LO INPUT, and HI SENSE to LO SENSE.
Current input (12A and 100mA) terminals.
12A terminal and LO terminal are used in current measurement of 1A and 12A measurement
ranges, and fuses inside the multimeter provide 15A protection limit at most to the current
through 12A terminal. 400mA terminal and LO terminal are used in current measurement of
measurement range from 10uA to 100mA, and current input fuses on the rear panel provide
500mA protection limit at most to the current through 500mA terminal.

Note:

In order to avoid fuse blowout and multimeter damage, please use current input terminals as
indicated as follows.
1) 12A and 100mA input terminals are not allowed to be connected to the current measurement
loops at the same time.
2) If the valid values of currents to be measured AC+DC are from 100mA to 12A, and only 12A
and LO terminals are allowed to be used in measurement.
3) When measuring current, please correctly select the current input terminals according to the
predicted current size before connecting multimeter power.
4) Current input to 12A terminal can not exceed 15A at most, otherwise fuses inside multimeter
will be blown out; current input to 100mA terminal can not exceed 500mA at most, otherwise
current input fuses on rear panel will be blown out.

IEC measurement category II overvoltage protection.

In order to avoid electric shock, the product provides overvoltage protection in connection of the
mains which meet the following two conditions at the same time;
1. HI and LO input terminals are connected to mains under the measurement category II
conditions (as described in the following).
2. The maximum line voltage of mains is 300VAC.

Warning: IEC measurement category II includes electric devices connected to mains through an
outlet on branch circuit. Such devices include most small appliances, test equipment, and other
devices that plug into branch outlets.

Содержание ET3260

Страница 1: ...1 ET3260 6 1 2 Digit Multimeter ET3260A 6 1 2 Digit Multimeter User Manual Hangzhou Zhongchuang Electron Co Ltd...

Страница 2: ...tly use leading wire suite Please do not use damaged or worn out leading wire suite which may cause instrument damage or personal injury During probe using the fingers should be kept behind finger pro...

Страница 3: ...t of measurement range from 10uA to 100mA and current input fuses on the rear panel provide 500mA protection limit at most to the current through 500mA terminal Note In order to avoid fuse blowout and...

Страница 4: ...measure circuits where transient overvoltages could exceed this level Environmental Precautions The product complies with requirements indicated in WEEE Directive 2002 96 EC The affixed product label...

Страница 5: ...two parameters in one input signal such as display AC voltage value and AC frequency value during AC voltage measurement Conduct remote operation through interfaces of IEEE 488 ET3260 RS 232 LAN and U...

Страница 6: ...nce measurement 22 2 1 9 Continuity measurement 23 2 1 10 Diode measurement 23 2 2 Usage of Sensor 24 2 2 1 Thermal resistance sensor 24 3 Characteristic and Function 24 3 1 Measurement Configuration...

Страница 7: ...tive force error 39 5 1 2 DCV load error 39 5 1 3 DCI load error 40 5 1 4 Noise control 40 5 1 5 Resistance measurement considerations 42 5 2 AC Measurement Considerations 42 5 2 1 True virtual value...

Страница 8: ...This 6 1 2 digit multimeter provides a 480 320 TFT display and the equipped display can indicate current working status F1 F5 software define area measured value etc No Introduction 1 HI and LO sense...

Страница 9: ...esistance 4 wire frequency period on off diode sensor capacitance and ratio among which resistance 2 wire and resistance 4 wire share one key frequency and period share one and on off and diode share...

Страница 10: ...et display system information etc Redefine function 1 3 Rear Panel No Introduction 1 Power outlet 2 Power fuses 3 Power switch 4 Voltage selector 5 VCM output voltmeter completes measurement output 6...

Страница 11: ...power voltage selector installed on multimeter rear panel according to your power supply voltage 2 Connect the multimeter to AC power with power line supplied by accessories 2 Start multimeter Turn o...

Страница 12: ...during AC voltage measurement with major display being voltage measurement value and auxiliary display frequency measurement value Statistic data display interface Statistic Data Display Description U...

Страница 13: ...er of measurement points the first record value is 0 and increases by 1 every time gain a new measurement value ordinate axis represents measurement value 1 7 Measurement Wiring Note The main input LO...

Страница 14: ...0mA input terminals are not allowed to be connected to the current measurement loops at the same time Press to select DC current measurement function 1A 12A measurement is as shown in the following fi...

Страница 15: ...uct Basic Measurement 2 1 1 DC voltage measurement Basic information Measurement range 100mV 1V 10V 100V and 1000V Maximum resolution 100nV Operation description Press to select frequency measurement...

Страница 16: ...switch modes and refer to the chapter of User interface for details 6 If necessary press to configure the following measurement parameters integral time input impedance and zero calibration method and...

Страница 17: ...chapter of Measurement configuration for details 2 1 3 AC voltage measurement Basic information Measurement range 100mV 1V 10V 100V and 750V AC technology true virtual value measurement and AC coupli...

Страница 18: ...and high current measurement with different wiring methods Low current measurement range 10uA 100uA 1mA 10mA and 100mA High current measurement range 1A and 12A Maximum resolution 10pA Operation descr...

Страница 19: ...isplayed value refers to measurement current value and auxiliary displayed value refers to frequency measurement value of the current input signal 2 Connect test lines as indicated in the chapter of 1...

Страница 20: ...ement wiring 3 Select measurement range as required and may select automatic measurement or press the softkeys of Measurement range and Measurement range to select fixed measurement range which have s...

Страница 21: ...information Frequency measurement band 3Hz to 300 kHz Period measurement band 3 3us to 0 33s Measurement range 100mV 1V 10V 100V and 750V Input signal range 100mVAC to 750VAC Technology equal precisi...

Страница 22: ...to measurement capacitance value and auxiliary displayed value refers to the ratio between the measurement value and full measurement range 3 Connect test lines as indicated in the chapter of 1 4 Mea...

Страница 23: ...the beeper threshold display resistance value 2 Connect test lines as indicated in the chapter of 1 4 Measurement wiring 3 Press the softkey of Setup to set beeper threshold when the measured resista...

Страница 24: ...as is required press the softkeys of Graduation mark and Graduation mark or keys of and 4 Press the softkey of Relative to record the current value after which display the difference between actual me...

Страница 25: ...t a continuous individual tone Threshold resistance value can be set as any number from 1 to 1000 which can only be adjusted on the front panel Threshold resistance is stored in volatile memory and af...

Страница 26: ...power period is as follows Resolution Integral Time High speed 4 1 2 0 02 PLC Low speed 4 1 2 0 1PLC High speed 5 1 2 0 2PLC Low speed 5 1 2 1PLC High speed 6 1 2 2PLC Mid speed 6 1 2 10PLC Low speed...

Страница 27: ...integral times through keys of and and then press 3 1 6 Switch between front rear input terminals All the measurement conducted by input terminal can also be conducted by input terminal on rear termin...

Страница 28: ...ver measurement range appearing on front panel During frequency and period measurement multimeter only uses one measurement range to include all the input signals from 3Hz to 300 kHz During continuity...

Страница 29: ...e between input signal and existing relative value the two are both converted to dBm dB reading in dBm relative value in dBm The following table shows possible combination between math operation funct...

Страница 30: ...elative value to 0 Press softkeys of On Off to start or stop relative operation and after starting Relative indication can be seen on the right side of main interface After setting press the softkey o...

Страница 31: ...key of Default to reset the Relative value to 0 After setting press the softkey of Exit to go back to the former layer interface 3 2 4 Limit test Limit test function can be used to test whether it exc...

Страница 32: ...or lower limit with Excessively higher meaning exceed the specified upper limit and Excessively lower meaning exceed the specified lower limit When status switches from Pass to Excessively higher or...

Страница 33: ...eading on screen Single trigger When the trigger source is set as auto trigger the Single indicator in status column will light on and multimeter will capture one reading or readings in specified numb...

Страница 34: ...ives trigger signal it will read readings in specified number The selected samples are stored in volatile memory and multimeter will set the samples as 1 after cutting off power or resetting remote in...

Страница 35: ...ve the record into USB drive When the USB interface on front panel is connected to USB drive it can save history data to the measRecords folder under USB drive s root directory with the measurement da...

Страница 36: ...start or stop beeper output Press the softkey of Recover to restore to factory setting 3 5 2 Command set ET3260 and 3260A are compatible with command sets of Agilent 34401A Agilent 34410A and Fluke45...

Страница 37: ...y copy the software update package to this folder and plug the USB driver into the USB interface on multimeter front panel Press the key of System update after all the preparations to update the syste...

Страница 38: ...parameters to set multimeter and connect multimeter to computer USB terminal Pay attention to that computer might need several seconds to identify and establish connection with multimeter 4 4 LAN Inte...

Страница 39: ...ltage measurement And the best connecting method is copper copper crimping for the multimeter input terminal is made from red copper The following table lists common thermal voltage during heterogeneo...

Страница 40: ...ou should select multimeter s next higher measurement range to reduce the error to an acceptable level The following table lists multimeter internal resistances under different measurement ranges Meas...

Страница 41: ...uch as possible Test lead looseness or vibration may cause induction error voltage Fasten test leads when operating near magnetic field and if possible try to use magnetic shielding material or keep a...

Страница 42: ...lead and contact resistance and is generally used for automation test under the situation where there are resistance long cable or plentiful connections or switches between multimeter and equipment u...

Страница 43: ...low AC signal of high DC offset during measurement AC coupling true virtual value measurement is quite useful For example it is a common situation of measuring AC ripple wave existing in DC power How...

Страница 44: ...g 0 03 of the measurement range 5 2 3 ACV load error When using AC voltage measurement function multimeter input impedance is 1M with resistance being connected to 100pF capacitance in paralleled way...

Страница 45: ...r pre heating and in using Auto Zero function and the 24 hour indicator is relative to calibration standard and test standard complies with GB T 13978 2008 Digit Multimeter 6 1 DC Characteristics 6 1...

Страница 46: ...ranges other measurement ranges are all 1k wire Current source accuracy 3 On off test On off threshold 1 1000 which can be set by users Test current 1mA 3 Response time 300 samples second Display res...

Страница 47: ...0 200 1500 200 50 20 Resistance 4 10 10u 10mA 30 30 80 40 100 40 6 5 100 100u 1mA 30 30 80 40 100 40 6 5 1k 1m 1mA 20 5 70 10 100 10 6 1 10k 10m 100uA 20 5 70 10 100 10 6 1 100k 100m 50uA 20 5 70 10 1...

Страница 48: ...or is 0 13 of the measurement range Maximum input 750Vrms Input impedance 1 M 1 shunt capacitance 100pF Measurement method AC coupling virtual value response Maximum DC offset not lower than 1000V AC...

Страница 49: ...0 12 0 05 0 011 0 005 50k 100kHz 4 0 55 0 08 0 60 0 08 0 60 0 08 0 060 0 008 100k 300kHz 3 4 00 0 50 4 00 0 50 4 00 0 50 0 20 0 02 AC current 100uA 100 pA 3Hz 5 Hz 1 0 0 04 1 0 0 04 1 0 0 04 0 100 0 0...

Страница 50: ...eristics Gate time can be set as 1s 100ms and 10ms or a wider range Stability refers that after DC offset voltage changes the time to reach stable and accurate measurement result is not longer than 1...

Страница 51: ...nF 10uA 10pF 2sps 1 0 3 0 05 0 01 100nF 100uA 100pF 2sps 1 0 3 0 01 0 01 1uF 1mA 1nF 2sps 1 0 3 0 01 0 01 10uF 1mA 10nF 2sps 1 0 3 0 01 0 01 100uF 1mA 0 1uF 2sps 1 0 3 0 01 0 01 1000uF 10mA 1uF 2sps 1...

Отзывы: