Hangzhou Zhongchuang Electronics Co., Ltd.
15
Connect the instrument to the PC:
1. Locate the USB driver software in the CD.
2. Use the Mini-USB cable to connect the instrument to the USB port of the PC,
shown in Figure 22, press and hold the power key to start the instrument.
3. Install the USB drive;
4. After the installation is completed, check the serial number in Windows’ Device
Manager.
Figure 22
Flow Control
:
None
Configuration of virtual serial port:
ET1080 Series employ the serial communication parameters with variable Baud
rates:
Baud rate: 9600 or others;
Data bits: 8
Check: None
Stop bit: 1
Flow Control: None
Remote Control State:
When ET1080 Series receives the remote operation state instruction from the host,
the instrument automatically enters into the remote control state. "RMT" is displayed on
the LCD screen to show the entry into the remote control state. To exit the remote control
state, send the “SYSTem: LOCal” command.
Command Protocol:
ET1080 Series handheld LCR uses SCPI command set to transfer control command
and return query information and data with string. The terminator specified by the protocal
shows the end of a command line or information enquiry line.
Using SCPI command set enables the interaction control of PC over the instrument
by programming. The command format meets the standard which is easy to understand
and use.
Public command: the command applied universally to various kinds of instrument
defined by the public command IEEE488, the public command starts with *, ET1080
series supports only * IDN?