background image

 

The material contained in this document is the property of Electronics & Innovation Ltd., it is 
subject to change without notice.  
 

August 2012                                                                                                                             

Revision B 

 

13 

c) 

Spectrum Analyzer.   

d) 

Attenuator (30dB) 

e) 

Coupler (30 dB) 

 

2.) 

Connect the Equipment as shown in Figure 4-2, then proceed as follows: 

 
a) 

Adjust the signal generator to a CW center frequency of 10 kHz, for an 
indicated output of 400 watts on the power meter. 

b) 

Using  the  spectrum  analyzer,  check  that  the  level  of  the  carrier 
harmonics  is  less  than  <  -21  dBc  with  respect  to  the  carrier  while 
manually  scanning  the  frequency  band  of  10  KHz  to  10  MHz.  An 
indicated  power  output  of  250W  should  be  maintained  during  this 
operation. 

 
 
4.5     PACKAGING FOR RESHIPMENT 
 
In the event of the equipment being returned for servicing it should be packed in 
the original shipping carton and packing material. If this is not available, wrap the 
instrument  in  heavy  paper  or  plastic  and  place  in  a  rigid  outer  box  of  wood, 
fiberboard  or  very  strong  corrugated  cardboard.  Use  ample  soft  packing  to 
prevent  movement.  Provide  additional  support  for  projecting  parts  to  relieve 
these  of  unnecessary  shock.  Close  the  carton  securely  and  seal  with  durable 
tape. Mark the shipping container FRAGILE to ensure careful handling. 

 

 

 
Chapter 5 Safety: 
 

Do  not  attempt  to  operate  this  unit  with  the  cover  removed.  High  AC  and  DC 
voltages are present. The cover protects against electrical shock due to AC line 
voltages,  high  DC  and  RF  fields.  Further  the  cover  provides  part  of  the  cooling 
system  design.  Components,  specifically  on  the  RF  driver  board  are  prone  to 
over-heat and eventual failure if the unit is operated without the cover in place. 
 
Ensure that the load is connected to the output prior to connecting the RF input 
to  the  unit.  This  will  prevent  high  voltages  being  present  and  exposed  at  the 
output connector.  
 
Only use the AC cord provided or equivalent.  
 
Ensure that the mains outlet is properly grounded. 
 

Содержание 2400L

Страница 1: ...out notice August 2012 Revision B 1 2400L Broadband Power Amplifier HIGH RF VOLTAGES MAY BE PRESENT AT THE OUTPUT OF THIS UNIT All operating personnel should use extreme caution in handling these volt...

Страница 2: ...lly operational such that it performs according to the original specifications free of charge to the original purchaser Should warranty service be required the unit must be returned to E I freight cos...

Страница 3: ...SB pulse and any complex modulation signals The amplifier has 55 dB gain it is unconditionally stable and will not oscillate even with combinations of mismatched source and load impedance It is protec...

Страница 4: ...lts RMS for all output load conditions without damage POWER OUTPUT METER True RMS power detection 3 of full scale accuracy POWER REQUIREMENTS 100 240 VAC 47 63 Hz SIZE H x W x D 13 x 16 75 x 20 inches...

Страница 5: ...5 times the level of the input signal required to achieve maximum output Input voltages in excess of 2 volts peak may permanently damage the instrument iii Connect the output via a 50 ohm coaxial lead...

Страница 6: ...suitable for connection to a PC or host system The communication protocol is extremely simple to facilitate readback and control with readily available terminal programs such as Hyperterm The RS 232...

Страница 7: ...further amplified signal appearing at the drain of Q1 is coupled to the input of Q2 This is transformed to 50 and fed to the driver output BNC port The driver output signal is fed through a length of...

Страница 8: ...bject to change without notice August 2012 Revision B 8 There are four switch mode power supply units three provide a 42 VDC 24 ampere source The other provides 42VDC and 24 VDC outputs and provides a...

Страница 9: ...ail safe under all load conditions Damage can only be externally caused by the incorrect selection of the AC supply voltage or by an input signal in excess of the specified 1 volt rms equivalent to a...

Страница 10: ...or c Adjust the output level of the sweep generator so that a 50 ohm video detector connected at the output of the unit will not be damaged by excessive power output Reference section 4 4 1 for set up...

Страница 11: ...D151 e Attenuator 30 dB 500 Watts Bird Figure 4 1 Gain Measurement 2 Connect the equipment as shown in Figure 4 1 then proceed as follows a Set the oscilloscope to DC Time cm to Ext X and gain to l0mv...

Страница 12: ...full deflection Rotate the step attenuator CCW so that the output is reduced by 50 dB g Reconnect the 2400L into the test set up of Figure 4 1 h Place the 2400L power switch to the ON position i Obser...

Страница 13: ...ap the instrument in heavy paper or plastic and place in a rigid outer box of wood fiberboard or very strong corrugated cardboard Use ample soft packing to prevent movement Provide additional support...

Отзывы: