OPERATING MANUAL
EI-4000
DIGITAL WEIGH INDICATOR/TRANSMITTER
THIS MANUAL APPLIES TO EI-4000 FIRMWARE VERSION 6.31 &ABOVE
Revised: 2015-04-07
EAGLE MICROSYSTEMS, INC
.
366 CIRCLE OF PROGRESS
POTTSTOWN, PA 19464
EI-4000 MANUAL IHPN=000000
Страница 1: ...L EI 4000 DIGITAL WEIGH INDICATOR TRANSMITTER THIS MANUAL APPLIES TO EI 4000 FIRMWARE VERSION 6 31 ABOVE Revised 2015 04 07 EAGLE MICROSYSTEMS INC 366 CIRCLE OF PROGRESS POTTSTOWN PA 19464 EI 4000 MAN...
Страница 2: ...uts two additional relay contacts and two 4 20mA analog outputs EI 2000 IO This board mounts on the EI 2000S CPU board It furnishes eight additional relay contacts and four remote switch inputs Note I...
Страница 3: ...display at the top of this page Tare weight This is the weight of the container e g cylinder tank or barrel Remaining weight a k a net weight This is the weight of the product e g Chlorine that the c...
Страница 4: ...If it does not press the ZRO ZERO pushbutton If the ZRO pushbutton is not displayed access the calibration mode to zero the scale 2 Place container on the scale 3 Press the SELECT pushbutton to go to...
Страница 5: ...RE Typical application partially used chlorine cylinders Remaining tare gross weights are displayable 1 An empty scale should read zero If it does not press the ZRO ZERO pushbutton If the ZRO pushbutt...
Страница 6: ...DOWN pushbuttons to adjust the gross scale calibration When satisfied that the weight is correct press the ENT pushbutton to save the new gross weight value EXTENDED MAIN SCREEN DISPLAYS rate daily us...
Страница 7: ...eight display screen VIEWING ALL WEIGHT VALUES ON ONE SCREEN MAIN MENU SET LOG ALL ESC SETUP MENU CFG CAL RTC ESC G 0 0 R 150 0 R 75 0 G 0 0 If more than one channel is enabled a screen that displays...
Страница 8: ...Low rate set point activated CHx DAILY HIGH High daily set point activated CHx DAILY LOW Low daily set point activated CHx SLOW REFILL Slow refill time has expired DAILY LOG The daily log displays the...
Страница 9: ...he meter See the operating modes section of this manual for a detailed description of each one P010 lb kg display only This displays the engineering unit lb or kg selected by S1 3 The user can change...
Страница 10: ...b kg 1 DCS302 300 140 0 5 0 2 1 2 60 27 2 150 136 2 EDS400 400 180 0 5 0 2 1 2 60 27 2 150 136 3 DS750 750 340 0 5 0 2 1 2 60 27 2 150 136 4 WT3600 1 ch 4000 1810 5 2 1 10 1000 454 2000 907 5 WT3600 2...
Страница 11: ...l scale is 300lbs and auto zero is set to 0 5 the weight must be below 1 5lbs for the meter to automatically zero the scale P114 Push to Zero Auto zero PTZ AZ maximum range 2 5 10 20 50 or ALL When th...
Страница 12: ...High level remaining weight set point The set points have a 1 hysteresis based on the full scale and an activation delay of three seconds For example if the full scale is 300lbs the set point is 10lbs...
Страница 13: ...the load is inductive add a snubber RC circuit across the relay contact terminals to greatly extend the life of the relay contacts The relays on the EI 2000 CPU TB2 CH1 CH2 and EI 2000 EXP TB10 CH3 CH...
Страница 14: ...ll below 4mA 0 lb kg or rise above 20mA until hardware limits are reached YES will limit the analog output to the 4mA to 20mA range SERIAL INTERFACE PARAMETERS P600 AOUT Gr NXT LAST ADJ ESC P610 A FS...
Страница 15: ...ts the decimal point for the rate When the number cannot fit on the display the decimal point is shifted to accommodate the rate value and can be shifted to where no decimal point is displayed P910 R...
Страница 16: ...t point The relay assigned can be shared with other set point relays Choices are OFF CH1 CH2 CH3 CH4 K1 through K8 K11 and K12 P933 D HIGH daily high level set point Entering a zero disables this set...
Страница 17: ...s NO Scale inputs are not summed P991 SUM SEL sum selection RE Remaining weights are summed Gr Gross weights are summed P992 SUM EXT sum external EI 4000 YES Sum all active scales Up to eight scales a...
Страница 18: ...le four is not enabled the CH4 pushbutton PB4 is reassigned as an escape ESC back to the calibration menu After a channel is selected for calibration the zero span calibration screen is displayed The...
Страница 19: ...this must be taken into consideration when connecting the load to the analog outputs Refer to schematic above ANALOG 4mA and 20mA POINT ADJUSTMENT PARAMETERS Each analog output is calibrated to 4mA a...
Страница 20: ...g weight e g 3mV V at 4000lbs must be known this is the scale FULL SCALE entry that s written on the EI routing sheet Scale calibration dead zero one for each active channel The scale mV V dead zero v...
Страница 21: ...ENT SET NEW PASSWORD YES NO Power up functions are activated by first powering down the meter pressing the desired pushbutton combination and restoring power to the meter Restore all parameters and ca...
Страница 22: ...acter 0aH The numbers shown in the text boxes in blue if printed in color are character indexes for programming purposes RS232 Wiring to PC DB9 Connector 00000000001111111111222222222233333333334444 4...
Страница 23: ...RS 485 Wiring This is the wiring typical of an RS 485 interface...
Страница 24: ...ere 250VAC 5x20mm time lag fuse Do not use any other rating VOLTAGE SELECT SWITCH S2 The voltage selection switch located on the EI 2000S CPU selects the 115V or 230V range EI 4000 METER SELECTION HEA...
Страница 25: ...ading factory typical default This can be also performed by pressing the LOAD pushbutton in the scale full scale parameters P110 FACTORY DEFAULT Automatically loading default configuration calibration...
Страница 26: ...ich is 5V 5 2 S Positive signal Positive scale signal input 3 S Negative signal Negative scale signal input 4 E Negative excitation Negative power supply voltage return for the scale base which is 0 v...
Страница 27: ...the scale base which is 5V 5 2 S Positive signal Positive scale signal input 3 S Negative signal Negative scale signal input 4 E Negative excitation Negative power supply voltage return for the scale...
Страница 28: ...INSTALLATION SPARE PARTS LCD1602 EI 110392 CPU P N 110394 EI 2000 IO 110396 EI 2000 EXP 110395 Interconnecting ribbon cable 8 111026...
Страница 29: ...00645 WP1000 1 330 150 1 4 N RED GRN WHT BLK YEL 500736 or 500737 WP1000 2 661 5 300 1 4 N RED GRN WHT BLK YEL 500725 WP1000 3 1323 550 1 4 N RED GRN WHT BLK YEL 500722 DS750 750 340 2 1 N RED GRN WHT...
Страница 30: ...ight 3 lb kg R float 2 0x101e 0x201e used weight 4 lb kg R float 2 0x1018 0x2018 rate 1 dwt dt R float 2 0x101a 0x201a rate 2 dwt dt R float 2 0x101c 0x201c rate 3 dwt dt R float 2 0x101e 0x201e rate...
Страница 31: ...point 1 state 3 N A R uint16 1 0x1183 set point 1 state 4 N A R uint16 1 0x1184 set point 2 state 1 N A R uint16 1 Set point two state 1 ACTIVE and 0 INACTIVE 0x1185 set point 2 state 2 N A R uint16...
Страница 32: ...Notes...
Страница 33: ...INT MODE 1 P500 STOP BITS 1 OR 2 P703 SET POINT MODE 2 P501 PARITY EVEN ODD NONE P704 RTU NODE 1 32 P705 CHANNEL SETUP SCALE SEL P100 P200 P300 P400 FULL SCALE P110 P210 P310 P410 RESOLUTION P111 P211...