EI-1000 & 2000: Setup, Calibration and Troubleshooting.
Page 18 of 19
Note that the total amount in the tank is unknown at this point. As the 30 gallons
that were added are used the weight reading will drop to zero, then go negative.
When the water is at it’s lowest level, zero the scale.
Recommended:
After correct operation is verified, use the diagnostic mode to
write down the values of parameters 1.o, 1.c, 2.o, 2.c. These store the calibration
information. If the calibration or zero is altered these numbers can be reentered
to restore it.
Date( )
Date( ) Date( ) Date( )
Parameter
1.o (Channel 1 zero)
1.c (Channel 1 span)
2.o (Channel 2 zero)
2.c (Channel 2 span)
See the EI-1000 and EI-2000 Operating manual for more in depth information on
these instruments. We’ll be glad to email it to you if needed.
Contact Eagle Microsystems with any questions at 610-323-2250.