Page 7 of 12
1. Press and hold Button 3 for five seconds until the display reads
BEFORE
, then tap
Button 1 repeatedly until
DATE
displays.
2. Program the date in the following order:
a.
Year = YY
. Press Buttons 2 and 3 to set, and press Button 1 to advance.
b.
Month = MM
. Press Buttons 2 and 3 to set, and press Button 1 to advance.
c.
Day = DD.
Press Button 2 and 3 to set, and press Button 1 to complete.
3. Press Button 1 to set the next function.
If the unit returns to Normal mode, follow the steps outlined for setting each function.
Dispensing History Log (RECORD)
Your e-pill Med-Time XL can record when medications are actually dispensed (event
history log record). Every time medications are dispensed, an entry is created in the log
file.
The log entry records the actual time
(Time and Date)
when the previous 28-medication
dispensing events
(date and time for each medication event)
occurred. The event log
memory holds up to 28 events. You must set the current date (DATE) in order for this
feature to work.
1. Press and hold Button 3 for five seconds until the display reads
BEFORE
, then tap
Button 1 until
RECORD
displays.
2. The time that the last medication was taken automatically shows on the display
. To
view the previous 27 times (total number of records is 28), press Buttons 2 or 3.
3. Press Button 1 to set next function.
If the unit returns to Normal mode, follow the steps outlined for setting each function.
Perform Diagnostic Test (TEST)
The e-pill Med-Time XL has built-in functions to test the dispenser. To test the basic
functions of your e-pill Med-Time XL
perform these diagnostic tests.
Please run this test before you contact e-pill Medication Reminders or the manufacturer.
Test # 1: Alarm and Rotation (MOVING):
Test the normal dispensing functions.
1. Press and hold Button 3 for five seconds until the display reads
BEFORE
, then tap
Button 1 until
TEST
displays.
2. Press Button 3 to start diagnostic Test #1.
The word
MOVING
appears on the display
. The Medication Tray will rotate one sector,
the alarm sounds and
TAKE
appears on the display
.
3. Close the lid.
4. Turn Med-Time XL over to simulate medication dispensing. If everything is normal, the
alarm will stop and the word
TEST
will re-appear on the display.
If the word
ERROR
appears at the end of the test, contact e-pill Medication Reminders or
the manufacturer.