Press FVC, VC and MVV on the screen to review this patient’s testing report
respectively.
Press “PageUp” and “PageDw” button for more test results.
Note:
Once the data has been deleted, it cannot be recovered.
3.6 Test Validation
The goal of spirometry testing is to obtain accurate measurements of lung
function. A “Valid” test includes at least three technically “acceptable”
maneuvers. Once three acceptable curves are recorded, consistency or
repeatability of results is evaluated. When the subject records three acceptable
curves with repeatable results, the test is valid and can be concluded.
If all three tests cannot meet the requirement of test result validation, the
subject will need to attempt a few extra trials (No more than 8 trials in total) until
the correct data is collected.
Follow the steps below to repeat the test and replace the invalid test result:
Tap on the parameters that a patient wishes to repeat the test.
Follow the testing procedure accordingly.