42
@IST01530EEJ 01 917 5A0 19 041 1
Select the f igure to be
modif ied
with
the
key s .
Mo d ify the v alu e of
t he hi gh li ght ed f ig ur e
with keys .
New piping
►Existing piping
TIGHTNESS TEST
►Start test
Stabilization
Combustible gas
Test gas
EXISTING PIPING
Start test
►Stabilization
Piping volume
Combustible gas
Test gas
EXISTING PIPING
►Wait time
STABILIZATION
15 min
►Wait time
STABILIZATION
15
min
►Wait time
STABILIZATION
20
min
Start test
Stabilization
►Piping volume
Combustible gas
Test gas
EXISTING PIPING
►Volume up to 25dm
3
Volume calculated
EXISTING PIPING
Volume up to 25dm
3
►Volume calculated
EXISTING PIPING
►Start test
Stabilization
Piping volume
Combustible gas
Test gas
EXISTING PIPING
►Vtot
Volume measure
Add tube
Subtract tube
Zero volume
PIPING VOLUME
25.0 dm
3
►Vtot
Volume measure
Add tube
Subtract tube
Zero volume
PIPING VOLUME
02 .0 dm
3
5
Vtot
►Volume measure
Add tube
Subtract tube
Zero volume
PIPING VOLUME
25.0 dm
3
Take one syringe
(100 ml) of gas
From the piping
then press OK
VOLUME MEASURE
Inject the
syringe
contents in
the piping
VOLUME MEASURE
Pressure not
stable
Please wait ...
VOLUME MEASURE
Selected test is v alid for volume pipes up to 25 dm
3
(25 liters). It is not
required to know and enter the volume of the pipe since it is assumed to
be 25 dm
3
. In this way the leakage rate is actually rounded up and better
guarantee the v alidity of a “compliant” result. The test procedure is not
described in details because it is the same as the calculated volume
shown in the f low chart except f or the v olume size.
Automatically.
When
the
v olume
measurement procedure
is correctly terminated,
The 4400 shows the
m e a s u r e d
v a l u e ,
otherwise it requires to
repeat the gas injection
with the syringe.
1
2
3
4
5
6
7
8