Cerabar T PMP 135
Hauser
4
Electrical connection
Analog output connection
Supply voltage
12 to 30 VDC
Maximum 5%
Plug with 1/2” NPT, Pg 11, M 12 x 1 or cable
Residual ripple
Cable entry
1 Plug with 1/2” NPT or Pg 11
2 M 12 x 1 ;lug
3 Cable (rd = red, wh = white, gn = green
1 Plug with 1/2” NPT or Pg 11
2 M 12 x 1 ;lug
R: External load, e.g. relay, programmable logic controller, distributed control system
PNP switch connection
Supply voltage
Residual ripple
Cable entry
Current consumption
12 to 32 VDC
Without load, < 20 mA, with reverse polarity protection
Maximum 10%
Plug with 1/2” NPT, PG 11 or M 12 X 1
–
–
–
+
+
+
+
GND
–
1
2
3
rd
wh
gn
3
3
2
1
12 to 30 V
12 to 30 V
12 to 30 V
4 to 20 mA
4 to 20 mA
4 to 20 mA
2 –
1 +
–
–
+
+
2 –
1+
3+
R
R
1
2
18 to 32 V
3
2
1