User Manual EE260 Heated Humidity and Temperature Probe for Meteorological Applications
5 Setup and Adjustment
The EE260 is ready to use and does not require any configuration by the user. The factory setup of
EE260 corresponds to the type number ordered. Please refer to the data sheet at
. The user can change the factory setup with the help the EE-PCS Product Configuration Software,
the Modbus configuration adapter (order code HA011018) and the EE260 configuration cable (order
code HA011020).
Besides the measurand selection of the analogue and digital outputs and their scaling, the digital
interface settings may be changed. The RH and T measurement is offset and 2-point adjustable. It is
possible to enable or disable the sensing element and the probe heating (separated from each other;
factory setting for both: enabled) and the NAMUR error indication (factory setting: disabled). For the
purpose of pressure compensation, the barometric air pressure at the operating site may be set.
Please note:
The EE260 may not be connected to any additional power supply when using the Modbus
configuration adapter HA011018.
PC
Fig. 5
Configuration
5.1 EE-PCS Product Configuration Software
To use the software for performing adjustments and changes in settings, please proceed as follows:
1. Download the EE-PCS Product Configuration Software from
install it on the PC.
2. Connect the EE260 to the PC using the Modbus configuration adapter and the EE260 configuration
cable.
3. Start the EE-PCS software.
4. Follow the instructions on the EE-PCS opening page for scanning the ports and identifying the
connected device.
5. Click on the desired setup or adjustment mode from the main EE-PCS menu on the left and follow
the online instructions of the EE-PCS
5.2 Digital Interface RS485 with Modbus RTU Protocol
Item
Factory settings
Selectable values (via EE-PCS)
Baud rate
9 600
9 600, 19 200, 38 400, 57 600, 76 800, 115 200
Data bits
8
8
Parity
Even
None, odd, even
Stop bits
1
1, 2
Slave address
235
1...247
Tab. 2
Digital interface default settings
The recommended settings for multiple devices in a Modbus RTU network are 9600, 8, even, 1.
The EE260 represents 1 unit load on an RS485 network.
Device address, baud rate, parity and stop bits can be set via:
1. EE-PCS Product Configuration Software, the Modbus configuration adapter HA011018 and the
EE260 configuration cable HA011020.
The EE-PCS can be downloaded free of charge from
2. Modbus protocol in the register 60001 (0x00) and 60002 (0x01).
See Modbus Application Note AN0103 (available at
)
The serial number as ASCII-code is located in the register addresses 0x00...0x07 (16 bits per address).
The firmware version is located in the register address 0x08 (bits 15...8 = major release; bits 7...0 =
minor release). The sensor name is located in register address 0x09. The beforementioned registers can
be read out with function code 0x03 or 0x04.