D aeyoon S cale Industrial C o.,Ltd.
C hloride-M eter D Y -2501B
16
6. Maintenance
6-1. Storage and treatment of the sensor
The sensor is the most sensitive part of the Chloride-Meter. The following directions for care
should be observed to secure the reliability of the test results when re-using the sensor as
well as to ensure long-term use.
(1) Treatment of the sensor while measuring
- In case you have to stop the test temporarily, immerse the electrode in cleaning solution.
- Always keep the internal solution of the sensor at least 2/3 full.
(2) Storage
- When the sensor is not in use for more than one day, plug refill hole with cap and store
the sensor after put on protection cap.
(3) Re-using the sensor
Immerse the sensor in cleaning solution for an hour before you use.
6-2. Storage of meter
Item
Condition
Ambient Temperature
-5 ~ 50
℃
Avoid
places
where
sudden
changes
in
temperature coour which could cause freezing
or condensation.
Storing Temperature
-5 ~ 50
℃
Relative Humidity
30 ~ 90%
Environment
The place should be away from direct sunlight and free from
dust, corrosive gas, inflammable gases oil mists, steam, dripping
water or vibration, salty environments should be preferably be
avoided.
7. Warranty and A/S
7-1. Quality certification
(1) A quality certificate for this meter is included in the box.
(2) The warranty is valid for one year from the day of purchase. It covers defective parts of
the sensor which occurs within the warranty period. This warranty extends only to the original
retail purchaser.
However, it does not cover the misuse, abuse, negligence, modification of the sensor, or any
defects caused by a natural disaster. After the applicable period you must pay for A/S.