D aeyoon S cale Industrial C o.,Ltd.

C hloride-M eter D Y -2501B

16

6. Maintenance

6-1. Storage and treatment of the sensor

The sensor is the most sensitive part of the Chloride-Meter. The following directions for care

should be observed to secure the reliability of the test results when re-using the sensor as

well as to ensure long-term use.

(1) Treatment of the sensor while measuring

- In case you have to stop the test temporarily, immerse the electrode in cleaning solution.

- Always keep the internal solution of the sensor at least 2/3 full.

(2) Storage

- When the sensor is not in use for more than one day, plug refill hole with cap and store

the sensor after put on protection cap.

(3) Re-using the sensor

Immerse the sensor in cleaning solution for an hour before you use.

6-2. Storage of meter

Item

Condition

Ambient Temperature

-5 ~ 50

Avoid

places

where

sudden

changes

in

temperature coour which could cause freezing

or condensation.

Storing Temperature

-5 ~ 50

Relative Humidity

30 ~ 90%

Environment

The place should be away from direct sunlight and free from

dust, corrosive gas, inflammable gases oil mists, steam, dripping

water or vibration, salty environments should be preferably be

avoided.

7. Warranty and A/S

7-1. Quality certification

(1) A quality certificate for this meter is included in the box.

(2) The warranty is valid for one year from the day of purchase. It covers defective parts of

the sensor which occurs within the warranty period. This warranty extends only to the original

retail purchaser.

However, it does not cover the misuse, abuse, negligence, modification of the sensor, or any

defects caused by a natural disaster. After the applicable period you must pay for A/S.

Содержание DY-2501B

Страница 1: ...Chloride Meter DY 2501B Operation Manual DaeYoon Scale Industrial Co Ltd...

Страница 2: ...8 4 2 Calibration 4 2 1 Water soluble chloride Cl calibration 10 4 2 2 Acid soluble chloride Cl calibration 11 4 3 Measurement 4 3 1 Water soluble chloride Cl measurement 13 4 3 2 Acid soluble chlorid...

Страница 3: ...d between the sensing and reference electrodes is a measure of the activity of the reactive species As the activity of the species reacting at the sensing electrode varies so does the potential measur...

Страница 4: ...ered to CAUTION Denote operating procedures and maintenance must be followed to secure user safety and data accuracy WARNING Denote operating procedures and maintenance that if not properly followed m...

Страница 5: ...n one minute after the sensor has been inserted in the sample for the sensor to be stabilized After calibration is completed measure the standard solution to check the accuracy of the calibration If t...

Страница 6: ...is not in use for a long time in order to minimize waste of electricity 3 2 Specifications Measurement Object Concentration of chloride ion in hardened concrete Method ISE Ion Selective electrode Meas...

Страница 7: ...g mode MEASURE Start Calibration or Measurement Light LCD Light 1 Select first cal point 0 005 Cl Time setup PM 2 Select second cal point 0 05 Cl 3 Select third cal point 0 5 Cl 4 DATE Set date and ti...

Страница 8: ...on Solution water soluble 0 005 0 05 0 5 Cl 60ml 3Btl 6 Calibration Solution acid soluble 0 005 0 05 0 5 Cl 60ml 3Btl 7 Extraction Solution water soluble 40g 3Ea 8 Extraction Solution acid soluble 40g...

Страница 9: ...water soluble 0 005 0 05 0 5 Cl 500ml 2 2 Calibration Solution acid soluble 0 005 0 05 0 5 Cl 500ml 2 3 Internal Filling Solution 39ml 2 4 Calibration Solution water soluble 0 005 0 05 0 5 Cl 60ml 2...

Страница 10: ...points selecting option Power ON Select water soluble acid soluble calibration mode Enter to calibration mode Select Calibration Solution 0 005 Cl Water soluble Acid soluble Calibration 0 05 Cl Water...

Страница 11: ...PRINT Power ON Select Water Acid Soluble Cl NaCl Calibration Use cursor to choose measurement mode Measurement Start Measurement Complete Measuring Sample Saving data automatically Saving data automat...

Страница 12: ...ATC MTC mode ATC MTC ATC Automatic temperature compensation MTC Manual temperature compensation Load data from Memory MEMORY 10 10 1 1 7 PRINT Move up 10 record Move down 10 record 1 decrese 1 increas...

Страница 13: ...the sensor and the status of the calibration solution Follow the steps described in 1 and repeat calibration Allowed accuracy 10 of reading during measurement 0 5 0 045 0 055 Select calibration soluti...

Страница 14: ...ibration and troubleshooting Impurities on the surface of sensor affecting sensor sensitivity or contamination of the calibration solution may cause inferior calibration results Polish the surface of...

Страница 15: ...0g For water soluble chloride place 4g of the sample to be tested into 40g of the extraction solution shake for 10 minutes until sample sink in solution and start the measurement Refer to ASTM C 1218...

Страница 16: ...ple until the temperature gap diminishes below 5 before measuring 4 4 Other Functions 4 4 1 Time and date adjustment 25 0 ATC 1 Funct Mode 2004 11 09 Date Setting AM 11 25 10 4 DATE FUNCT Press FUNCT...

Страница 17: ...t Calibration solution has been contaminated Recalibration Change the calibration solution 2 Sensor Error The sensor is disconnected properly Check the connection of the connector 3 Sensor Unstable Th...

Страница 18: ...on for an hour before you use 6 2 Storage of meter Item Condition Ambient Temperature 5 50 Avoid places where sudden changes in temperature coour which could cause freezing or condensation Storing Tem...

Страница 19: ...and Counseling 1 If you encounter any problems or malfunctions while using the chloride Meter call the A S center 822 858 6284 5 or visit the store where you made your purchase Warning Never modify th...

Отзывы: