Instrument Maintenance Recommendations

 

Daily: 

Remove the orifice and clean thoroughly with brush and precision drill bit. Clean your indexer barrel with

 

cotton patches when hot. Piston should spin easily when placed into a clean barrel.

 

Weekly: 

Give good cleaning with brass brush. Use oven cleaner to clean the piston rod exposed metal and orifice.

 

Monthly or Long Term: 

Depending on your company policy, a calibration or calibration check may be needed on a 

monthly, quarterly or yearly basis. It is a recommended that a repetitive maintenance schedule be created for your 
instrument.

 

Clean unit: 

Clean the dust and dirt out of the electronics module with clean air every six months to one year, more 

often in dirty environments. Perform this with power removed from the system. Remove the orifice and clean 
thoroughly. Clean your barrel. Use oven cleaner to clean the piston rod assembly and orifice.

 

Check Mechanical Tolerances: 

All dimensions and tolerances per ASTM D1238 and ISO 1133-1. Piston tip 

diameter = 0.3730”/9.474mm +/

-

0.0003”/0.0076mm; piston tip length = 0.2500”/6.35mm +/

0.0050”/0.127mm. 

Assembled piston rod weighs 100 grams. GO/NO-GO gage works properly on the orifice. Orifice length = 

0.3150”/8mm +/

0.0010”/0.0254mm. Weights still weigh correctly. Digital Encoder calibration. Gages can 

be 

purchased from Dynisco Polymer Test.

 

Barrel Diameter: 

Once the barrel is extremely clean, all dimensional measurements are to be made at room 

temperature it can be checked using a bore gage. Gage can be purchased from Dynisco Polymer Test. At the time 
o

f manufacture, the barrel center bore measures 0.3760”/9.55mm +/

0.0002”/0.00508mm. All dimensions and 

tolerances per ASTM D1238 and ISO 1133-1.

 

Ask for help: 

Call Dynisco Polymer Test directly at (508) 541-9400 or visit http://www.dynisco.com/polytest-ser- 

vices and ask for technical support. To help us handle your questions as quickly as possible, have the following 
items ready before you call:

 

1.

 

Instrument name and model number (on back panel)

 

2.

 

Instrument serial number (on back panel)

 

3.

 

Current version of instrument firmware (Power up unit to see, version shown on 

“About” 

screen under 

“Maintenance

 

Tab”)

 

4.

 

Computer system make and model (if applicable)

 

5.

 

Current version of LaVA Suite software (if applicable)

 

Equipment Setup

 

Unpacking the Indexer

 

The Lab Melt Flow Indexer comes in a heavy duty, wood container. First, remove the container’s lid—

a bit tool 

should be within the crate packing documents on the outside of the container that can be inserted into an electric 
type drill to remove the lid and bracing screws. Several boxes may be packed inside along with the instrument; 
remove them and check that all boxes are received. For example, they are coded 1 of 5 or 3 of 5, indicating the total 
number is five.

 

For non-weight lift model instruments, the instrument is braced into the box with test weights and possibly 

accessories under the instrument in a foam pattern. Remove the instrument’s bracing. Minding that a base unit 

weights 45lbs/21.4Kg, either lift the unit out from the top of the container with two people or carefully lay the 
container on its side and slide the instrument out of the container.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

P/N:974179 

Rev: 0918    ECO: 49667

 

www.dynisco.com

 

 

 

Содержание LMFI 5500 Series

Страница 1: ...From lab to production providing a window into the process P N 974179 Rev 0918 ECO 49667 www dynisco com 1 LMFI 5500 Series Melt Indexer MonitoringYourConditionsForConsistentResults User s Guide...

Страница 2: ...ing items ready before you call 1 2 3 Instrument name and model number on back panel Instrument serial number on back panel Current version of instrument firmware Power up unit to see version shown on...

Страница 3: ...to change any information route method or process which is discussed or provided herein without notice Equipment or products made prior to or subsequent to the publication date of this manual may hav...

Страница 4: ...t Operation Buttons 16 Instrument Operation Miscellaneous Touch Areas 19 Instrument Operation Panel Overview 19 Test Calculations 20 Calculations Method A 20 Calculations for the Method A test are 20...

Страница 5: ...ICONS Expand Back Test Recipe Fault Codes Permissions Settings The LMFI 5500 Series Melt Indexer has a touch screen display in addition to ICON keys A keyboard is also displayed when settings and valu...

Страница 6: ...stering and third degree burns Wear approved clean thermally insulated gloves when handling these components Should injury occur immerse injured area in cold water and get immediate medical attention...

Страница 7: ...container and labelled with the following symbol Broken thermometers and their spilled mercury can be sent back to the manufacturer UPS will accept these packages provided they are labelled and the ma...

Страница 8: ...responding related ordinances for the use of work materials will have to be made available as a supplement to the operating instructions 3 4 5 6 7 8 Introduction A flow rate test is a measure of a pol...

Страница 9: ...enerally used for testing high flow rate polymers This is a no cuts test that is the most convenient for busy laboratories Measurement incc 10min Method D can be conducted only if material melt densit...

Страница 10: ...bore gage Gage can be purchased from Dynisco Polymer Test At the time of manufacture the barrel center bore measures 0 3760 9 55mm 0 0002 0 00508mm All dimensions and tolerances per ASTM D1238 and IS...

Страница 11: ...ity The bench top should also be able to withstand hot dies and tools being dropped on them If carpet is present on the floor near the LMFI carpet protection is necessary since a hot die hot material...

Страница 12: ...1 LMFI with encoder and auto cutoff Figure 1 shows LMFI with the encoder and auto cutoff option Figure 2 shows the LMFI base unit that has no options Figure2 LMFI base unit Rear View Power Switch Pow...

Страница 13: ...m Part No Qty Each Mirror LMI 5500 1196030 1 Barrel If worn and out of specification 4051 25A 1 Piston Assembly Tip Piston Weight Top 7051 72 1 Piston Tip 0051 41 1 Die Orifice Standard 8mm Height 005...

Страница 14: ...top left icon to expand descriptions User Groups User Groups Default User Password Access Rights OPERATOR Testing only Cannot modify programs that are stored in memory Canruntestsandchangetestvariable...

Страница 15: ...From lab to production providing a window into the process Instrument Operation Icons P N 974179 Rev 0918 ECO 49667 www dynisco com...

Страница 16: ...P N 974179 Rev 0918 ECO 49667 www dynisco com Etitkdjkdj Create Edit Recipe Use an existing recipe or create a new one then select RUN which will bring you to the Operation page...

Страница 17: ...system has 3 levels of default user types Push the top left ICON to expand descriptions The HOME screen below is the v indicates actual operation temperature P N 974179 Rev 0918 ECO 49667 www dynisco...

Страница 18: ...From lab to production providing a window into the process P N 974179 Rev 0918 ECO 49667 www dynisco com...

Страница 19: ...to production providing a window into the process From lab to production providing a window into the process Machine Settings Instrument Operation PanelOverview P N 974179 Rev 0918 ECO 49667 www dyni...

Страница 20: ...me to traverse the Method B distance L L Method B length of flag mm M is the mass in grams of material collected over the time TA in seconds MFR has units of g 10 min grams 10 minutes g cc is called t...

Страница 21: ...would land on the top of the die 25 4mm Recently Method B has become the more common test because it is simpler to run hands free after material load and more precise for routine analysis In addition...

Страница 22: ...cosity by J Reilly and P Limbach available from DYNISCO POLYMER TEST on request The Amount of Sample ASTM gives a recommendation of how much material to put in the barrel to perform a test However by...

Страница 23: ...the instant the material is packed into the barrel The following table shows an estimate of charge weight in GRAMS based on anticipated MFR g 10 min and melt density g cc Where the word PLUG appears i...

Страница 24: ...From lab to production providing a window into the process Create Edit Recipe xx P N 974179 Rev 0918 ECO 49667 www dynisco com...

Страница 25: ...From lab to production providing a window into the process P N 974179 Rev 0918 ECO 49667 www dynisco com...

Страница 26: ...From lab to production providing a window into the process System Configuration P N 974179 Rev 0918 ECO 49667 www dynisco com...

Страница 27: ...From lab to production providing a window into the process Setting Up a Test P N 974179 Rev 0918 ECO 49667 www dynisco com...

Страница 28: ...o achieve the desired packing force The packer must be enabled through the System Configuration screen Material is loaded as stated above with a funnel The funnel is removed The packer is pivoted arou...

Страница 29: ...temperature stabilization packing loading of material weight selection and placement encoder position if used for test and melt time for starting of test Once the test has begun follow the screen pro...

Страница 30: ...into the barrel This allows the piston plunger and the die to reheat before the next test With materials that are thermally stable less than 5 viscosity change over 1 2 hour of heat exposure we recom...

Страница 31: ...ou may wish to coat the barrel with a light machine oil to prevent rusting The oil will need to be cleaned with cleaning patches from the machine before accurate data can be obtained Warning Do not he...

Страница 32: ...ber is NIST Special Publication 260 To order 301 975 6776 Fax 301 948 3730 2 Mercury Hg spill kits Mercury Clean Up Spill Kits Mercon Products distributed by Fisher Scientific Unit 8 7551 Vantage Way...

Страница 33: ...gle MFR Tests Up to 15 MFR readings per Barrel Filling can be obtained Any test flag length up to 30 mm can be used User Defined test length and test spacing for single or multiple tests can be used A...

Страница 34: ...nto place calibration is rarely required If calibration needs to be checked due to suspected encoder damage then the following steps should be followed If damage to the arm is suspected then contact D...

Страница 35: ...nto the process 2 Press the selection for Calibration using the Settings icon at the lower left of the screen 3 Press the display to bring up the calibration screen for Temperature and the Encoder P N...

Страница 36: ...ion of encoder arm 11 12 13 14 15 Press the display o save to save both calibration locations Screen reverts back to the calibration screen Lower arm back into its locking holder Remove weight from pi...

Страница 37: ...fore the test sample time is to begin The Auto Cutter option is triggered from the LMI main PCB via an 8 pin connection to the PCB The Auto Cutter option cuts in a single 360 degree rotation and drops...

Отзывы: