850plus & 550plus
Operation & Treatment Instructions
21
7. Administer treatment until the treatment time expires, or press STOP to stop all
treatments in progress at any time.
Each of the therapies you can deliver with the Dynatron 850plus and 550plus offers its own
default settings. To make your treatment setup even more convenient, the Dynatron 850plus
and 550plus allow you to change the defaults for each modality to your own most common
treatment selections. See “Setting Defaults” later in this manual.
The following pages provide more detailed setup instructions for each individual modality.
Before administering a treatment to a patient, be sure to read all treatment instructions
as well as contraindications, warnings, and precautions for that modality provided later
in this manual. Also, consult published literature for recommended application of these
modalities.
Conductance Bar Graph
This feature for models having software greater than Rev. 1.01 only (see Page 2).
During setup of an interferential or premodulated treatment, a conductance graph is
automatically displayed in the Time display while you set the treatment intensity. The graph
allows you to see whether conductance is good for that treatment setup and if current is being
delivered efficiently through the electrodes. The display reverts to the Time display when
you press any other key after setting the intensity. You may again view this conductance
graph at any time during an interferential, premodulated, or microcurrent treatment by
pressing FUNCTION-TIME. The conductance bar graph is not available for combination
treatments.
The bar graph uses the eight segments in the top half of the display, with eight lighted
segments indicating best conductance, and one lighted segment indicating poorest
conductance. In the illustration below, eight of the eight segments are lighted indicating good
conductance.
POWER-INTENSITY
TIME
START
STOP
PAUSE
WIDTH
RATE
FREQ
TIME
SOUND
MICRO
HV 2
HV 1
4
3
2
1
FUNCTION
During an IFC, Premodulated, or Microcurrent treatment, press
the FUNCTION and the TIME keys simultaneously to see the
conductance graph. Press TIME to return to the treatment
time display.
The graph is automatically displayed when the intensity is
increased during treatment setup for IFC and Premod.
CH
Conductance Bar Graph
Eight lighted segments indicate conductance is good. If only
two or three segments are lighted, conductance is poor.
Conductance Bar Graph.
Содержание Dynatron 550 Plus
Страница 1: ...i SERVICE MANUAL Dynatron 850plus Dynatron 550plus ...
Страница 9: ...Dynatron 850plus 550plus 1 Section I Introduction ...
Страница 27: ...Dynatron 850plus 550plus Operation Treatment Instructions 19 Section II Operation and Treatment Instructions ...
Страница 77: ...Dynatron 850plus 550plus General Modality Information 69 Section III General Modality Information ...
Страница 106: ...Dynatron 850plus 550plus Ultrasound Technical Information 98 ...
Страница 110: ...Dynatron 850plus 550plus Ultrasound Technical Information 102 1 MHz 2 MHz 3 MHz 5 cm2 Head Near Field ...
Страница 111: ...Dynatron 850plus 550plus Ultrasound Technical Information 103 1 MHz 2 MHz 3 MHz 2 cm2 Head Near Field ...
Страница 120: ...Dynatron 850plus 550plus Ultrasound Technical Information 112 ...
Страница 121: ...Dynatron 850plus 550plus Service 113 Section VI Technical Information ...
Страница 130: ...Dynatron 850plus 550plus General Technical Information 122 ...
Страница 131: ...Dynatron 850plus 550plus Service 123 Section VII Service ...
Страница 135: ...Dynatron 850plus 550plus Schematics and QC Check Lists 127 Section VIII Schematics and QC Check Lists ...
Страница 139: ...131 ...
Страница 140: ...132 ...
Страница 141: ...133 ...
Страница 142: ...134 ...
Страница 143: ...135 ...
Страница 144: ...136 ...
Страница 145: ...137 ...
Страница 146: ...138 ...
Страница 147: ...139 ...
Страница 148: ...140 ...
Страница 149: ...141 ...
Страница 150: ...142 ...
Страница 151: ...143 ...
Страница 152: ...144 ...
Страница 153: ...145 ...