Installation & Features
16
Soundheads for the Dynatron 850plus come in 1 cm
2
, 2 cm
2
, 5 cm
2
, and 10 cm
2
sizes. The
2 cm
2
, 5 cm
2
, and 10 cm
2
soundhead operate at 1, 2, and 3 MHz; while the 1 cm
2
soundhead
operates at 2 and 3 MHz. The soundheads are waterproof allowing ultrasound (not
combination) therapy to be administered in water, if desired.
Do not drop the soundhead nor allow it to strike a hard surface. Do not place soundheads
into ice water. Do not allow soundheads to repeatedly reach maximum head temperature.
All of these conditions can cause damage to the crystal. Any such damage is not covered by
the warranty.
For new or replacement soundheads, contact Dynatronics or your authorized Dynatronics
representative.
Microcurrent Probes (optional accessory)
The Dynatron 850plus and Dynatron 550plus devices offer optional probes for delivering
hands-on microcurrent treatment. The microcurrent probes can only be used for microcurrent
therapy. It is not possible to select the probe output nor to use the microcurrent probe to
deliver a probe treatment for interferential, premodulated, Russian, biphasic, or high volt
modes.
The setup includes an active probe along with either a ground probe or a ground electrode.
Use the end of a cotton swab (such as a Q-Tip
®
) inserted into the ends of both the active
probe and the ground probes. Cut the end of the swab to a short length; the cotton must touch
the probe’s metal ring. Use water or a conductive electrolyte spray to wet the cotton swab
before treating.
Active Probe: To deliver the current to the patient through the active probe, press and
release the button on the probe. Press and release the button again to stop the current. The
active probe should touch the patient’s skin at the treatment site, and the ground probe should
touch the patient’s skin elsewhere. This completes the circuit and delivers current to the
patient. You do not need to hold the button down. Once you have pressed and released the
button, the current is delivered until you press and release the button again to stop the current.
While you are delivering
current, the treatment
timer counts up in
seconds from zero. When
the current is stopped, the
timer returns to zero.
Press the button again to
commence the next time
sequence. Continue in
this way until treatment is
completed.
Ground Probe: The
ground probe is attached
directly to the banana pin
connector on the probe
cable. This probe is used to complete the circuit to allow flow of current through patient
tissue. The ground probe should touch the patient’s skin at any location away from the
treatment point. As an alternative, you may also use a ground electrode as explained below,
Press and release button
to start and to stop
current delivery during
treatment.
Cut the end off a cotton swab and insert it into
the tip of the cone. Saturate the cotton with
water or an electrolyte spray before treating.
Microcurrent Active Probe (optional accessory)
Содержание Dynatron 550 Plus
Страница 1: ...i SERVICE MANUAL Dynatron 850plus Dynatron 550plus ...
Страница 9: ...Dynatron 850plus 550plus 1 Section I Introduction ...
Страница 27: ...Dynatron 850plus 550plus Operation Treatment Instructions 19 Section II Operation and Treatment Instructions ...
Страница 77: ...Dynatron 850plus 550plus General Modality Information 69 Section III General Modality Information ...
Страница 106: ...Dynatron 850plus 550plus Ultrasound Technical Information 98 ...
Страница 110: ...Dynatron 850plus 550plus Ultrasound Technical Information 102 1 MHz 2 MHz 3 MHz 5 cm2 Head Near Field ...
Страница 111: ...Dynatron 850plus 550plus Ultrasound Technical Information 103 1 MHz 2 MHz 3 MHz 2 cm2 Head Near Field ...
Страница 120: ...Dynatron 850plus 550plus Ultrasound Technical Information 112 ...
Страница 121: ...Dynatron 850plus 550plus Service 113 Section VI Technical Information ...
Страница 130: ...Dynatron 850plus 550plus General Technical Information 122 ...
Страница 131: ...Dynatron 850plus 550plus Service 123 Section VII Service ...
Страница 135: ...Dynatron 850plus 550plus Schematics and QC Check Lists 127 Section VIII Schematics and QC Check Lists ...
Страница 139: ...131 ...
Страница 140: ...132 ...
Страница 141: ...133 ...
Страница 142: ...134 ...
Страница 143: ...135 ...
Страница 144: ...136 ...
Страница 145: ...137 ...
Страница 146: ...138 ...
Страница 147: ...139 ...
Страница 148: ...140 ...
Страница 149: ...141 ...
Страница 150: ...142 ...
Страница 151: ...143 ...
Страница 152: ...144 ...
Страница 153: ...145 ...