3. Installation
3-1. Power On/off and Door Release
3-1-1. Power On/off
Action
1. Connect the AC Power cord at the power socket on the back
of the instrument.
2. Turn on the instrument by pressing a switch on the back
of the instrument.
3. Press the ‘Door’ button to open the door.
3-1-2. Door Release
Action
1. For opening the door, press the [DOOR] button.
¾
The door is automatically opened after
completion of spinning operation with
beeping sound.
¾
When you close the door, close it until
hearing clank shut.
☞
9
The door is not opened while the instrument is running.
9
If the door is opened, the instrument could not operate even with pressing the ‘Start’
button.
9
Power Failure: If there is any power failure during the operation, door is not opened
with ‘Door’ button. Door can be opened only when the operation is completely
stopped and the power is on again. If you want to open the door at the power failure,
please refer to ‘4-6. Emergency Door Open’.
3-2. Rotor coupling and disassembling
Action
1.
Before coupling a rotor, clean the motor shaft and chamber with soft dry
towel.
2.
Mount a proper rotor into the motor shaft.
Fixed Angle Rotor