5-5. Transportation of the instrument
1. If you need to move or ship the instrument, be cautious to protect the motor shaft from any
physical impact or turbulence.
2. Do not mount a rotor in any cases of movement. Fill inside the chamber with proper materials
to keep the motor shaft on place and not to be influenced by physical pressure.
6. Trouble Shooting
6-1. Check list
Symptom
Check list
Power failure
Connect the AC Power cord and make sure that the line is completely
connected between the instrument and power outlet. Check the
power switch is turned on. (Please refer to 3-1. Power On/Off)
Can’t be started
If the door is not closed completely, the instrument can’t run.
Check the Door LED on the display window and close the door
completely.
Can’t open the door
If the power is out, check the main fuse for the laboratory to supply
the power. If it is not solved in shortly, open the door with spikes for
safety of sample. (Please refer to 4-6. Emergency Door Open)
Can’t close the door
Remove the dirt at the door latch and then close the door completely
again. If the door seems not being closed by mechanical reason,
please contact our service team.
Please check the balanced status of both the table and the
instrument
Noise and vibration during
running
Please re-check the coupling status of the following three matches to
minimize the noise
1. the balanced way of coupling of the rotor into the motor
shaft
2. the completeness of fixing of the Rotor Locking Nut on the
rotor
3. the matching status of Rotor Lid with the rotor