DNAmaster Quick Manual 

 

Version 1.0   

   

 

 

 

 

 

 

 

 

Page

 1 

The Instrument 

 

 

 

 

 

 

On/Off button 

Blue backlight- Standby mode 

Green backlight- Operating 

 

Read button 

The Read button on the unit can use as an alternative way to take the data during measurement. 

 

Touch screen 

DNAmaster’s LCD display is pressure sensing which can be used even with gloves on. For best 

performance, fingertip or stylus is preferable. 

Touch screen 

display 

Take 

Measurement 

on/off button 

Содержание HALO DNAmaster

Страница 1: ...Model HALO DNAmaster Bio spectrophotometer HALO DNAmaster Bio spectrophotometer Operation Manual Revision D01 201203 ...

Страница 2: ...ng measures Increase separation between the instrument and radio TV receiver Connect the instrument to an outlet on a circuit different from that to which the radio TV receiver is connected 2 Possible Electromagnetic Wave Interference Affecting This Instrument If this instrument is used near an intense electromagnetic source interference noise may be given to the instrument to incur an adverse eff...

Страница 3: ...rate according to the specifications given in the instruction manual provided it is used in accordance with the instructions described in the manual 1 Scope of Warranty a Any parts which prove to be defective in design or workmanship during the warranty period will be repaired without charge b A substitute part may be used for repair or replacement with an equivalent product may be made instead of...

Страница 4: ... to use of hardware software or spare parts other than supplied by us f Failure due to improper handling or maintenance by the user g Failure due to maintenance or repair by a service agent not approved or authorized by us h After disposal of this instrument or after its resale without our approval i Failure due to relocation or transport after initial installation j Failure due to disassembly mod...

Страница 5: ...om this warranty Any other implied warranties of merchantability and fitness for a particular purpose are not included in this warranty No liability is assumed for direct or indirect damages arising out of explicit or implied warranties b Oral or written information or advice given by our dealers distributors agents or employees without our express permission shall not create a warranty or in any ...

Страница 6: ...ed and discarded as instructed by the respective suppliers 2 Notice on Instruction Manuals 1 Information contained in the instruction manuals furnished with the instrument is subjected to change without notice for product improvement 2 This manual is copyrighted by Dynamica with all rights reserved 3 No part of this manual may be reproduced or transmitted in any form or by any means without our ex...

Страница 7: ...t The hazard warnings which appear on the warning labels on the product or in the manual have one of the following alert headings consisting of an alert symbol and a signal word DANGER WARNING or CAUTION DANGER Indicates an imminently hazardous situation that if not avoided will result in death or serious injury It does not apply to this equipment WARNING Indicates a potentially hazardous situatio...

Страница 8: ...ly related to the safety of a person Attention It is used to present warnings which are not directly related to personal injury hazards It is used to indicate prevention against damage to the equipment Note This is used to indicate instructions that enable you to operate the equipment accurately and perform accurate measurements ...

Страница 9: ... you do not follow the instructions in this manual an inaccurate analysis may result or bodily injury may occur Because of danger please make sure not to modify or alter the product make sure not to use unspecified parts and make sure not to operate the equipment by removing defeating the safety device s When using chemicals please make sure to ventilate the room well If there is not enough ventil...

Страница 10: ...installer Installation of this instrument shall be carried out by or under supervisions of qualified service personnel of Dynamica or its authorized service agent When relocation of this instrument becomes necessary after initial installation delivery please notify your local Dynamica sales representative or nearest Dynamica service office Technical support service for this instrument is available...

Страница 11: ...armful to your health WARNING Eye Injury from Organic Solvents Handling Organic Solvents Please wear protective glasses when using organic solvents If the organic solvent should get into your eye flush your eye immediately under running water for at least 5 minutes while keeping your eyelids open See a physician for appropriate treatment ...

Страница 12: ...ed WARNING Electrical Shock Electrical Shock from Contact High voltage is used inside of the equipment Never remove the cover except by a service person or as instructed in the section for lamp replacement When cleaning the inside of the cell compartment be sure to turn off the unit CAUTION High Temperature Burns from high temperatures The lamp will become very hot during operation please make sur...

Страница 13: ... hand or hair caught Please assure your safety before operation CAUTION Fatigue due to Prolonged Work Data Processing Viewing the display in your work can cause eye and physical fatigue if you continue to work in the same posture for extended periods When working with the display for a prolonged period for your health make sure to take breaks for 10 to 15 minutes every hour in order to rest your e...

Страница 14: ... NOTE Precautions regarding Accuracy and Precision of Measurements Do a routine check prior to taking measurements In order to obtain accurate measuring results please perform tests as instructed following the instructions in Chapter 3 ...

Страница 15: ...ve grounding may not only cause lower resistance against noise from the outside but it can also cause the Mass Spectrometer and personal computer to generate static electricity which may involve the danger of electrical shocks A high voltage circuit is used inside the Spectrophotometer Do not open the covers when this circuit is operational because of the danger from electrical shock Fire Extingui...

Страница 16: ...paint such as from organic solvents particularly benzene thinner etc 5 Other general notes a Avoid direct sunlight otherwise optical performance might deteriorate or the housing might become discolored Avoid installation by a window if at all possible b Vibrations or shocks strong enough to be felt by the human body must not be transmitted to the instrument otherwise the fine adjustment mechanism ...

Страница 17: ...Operating Read button The Read button on the unit can use as an alternative way to take the data during measurement Touch screen DNAmaster s LCD display is pressure sensing which can be used even with gloves on For best performance fingertip or stylus is preferable Touch screen display Take Measurement on off button ...

Страница 18: ...hotometric Repeatability 0 005 OD Stray light 0 5 T DNA detection limit 20ng µl Minimum sample volume with Ultramicro Cell Depends on the cap used 0 5µl Start up melodies Selectable from 7 types and mute Energy save mode Yes Memory storage Internal or SD card Data storage number 40 000 internal Unlimited external Manual program storage number 10 for each function Dimension W x D x H mm 300 x 300 x...

Страница 19: ... go to next screen DNA RNA Nuclei acid analysis DNA RNA DYE Dye incorporation PROTEIN Protein determination PROTEIN UV Protein UV UV dye method OD600 Bacterial cell culture measurement SPECTRO Traditional spectrophotometer function FILE Saved method and data SETTINGS User parameter setting ...

Страница 20: ... for blank or reference sample before measurement 4 READ for measurement 5 PRN for result printing Note Press ZERO key after new setting this will be used for all subsequent samples SET Parameter setting Input 1 Factor 2 320nm background correction on off 3 Pathlength 4 Dilution Factor 5 Unit 6 Sample Name Press OK to next step or Save to store the parameters Select between different sample number...

Страница 21: ... detail setting 2 LOAD for reload saved parameter 3 ZERO for blank before measurement 4 READ for measurement 5 PRN for result printing SET Parameter setting Input 1 Factor 2 background correction on off 3 Pathlength 4 Dilution Factor 5 Unit 6 Sample Name Press OK to next step or Save to store the parameters Select between different sample number ...

Страница 22: ...type Input 1 Dye Type For custom dye Input 1 Dye Abs Max for maximum absorbance wavelength of the custom dye 2 Dye Ext Coefficient for dye dependent extinction coefficient 3 Dye Correction for dye dependent correction factor at 280 nm Press OK to next step or Back for return to previous screen ...

Страница 23: ...nce sample before measurement 4 READ for measurement 5 PRN for result printing Note Press ZERO key after new setting this will be used for all subsequent samples SET Parameter setting Input 1 Pathlength 2 Calibration mode Standard concentration or input K factor 3 Number of standards 4 Dilution factor 5 Unit selection 6 Sample Name Press NEXT for Standard concentration or K factor input OK to next...

Страница 24: ...ration of standard used Press OK to next step or Back to previous screen For standard measurement For Standard curve setup Press 1 ZERO for blank before measurement 2 READ for measurement 3 PRN for result printing Note Press ZERO key after new setting this will be used for all subsequent samples ...

Страница 25: ...eturn to previous screen PROTEIN SAMPLE MEASUREMENT Press 1 ZERO for blank or reference sample before measurement 2 READ for measurement 3 PRN for result printing Note Press ZERO key after new setting this will be used for all subsequent samples Change to standard curve display Select between different sample number ...

Страница 26: ...ple before measurement 4 READ for measurement 5 PRN for result printing Note Press ZERO key after new setting this will be used for all subsequent samples SET Parameter setting Input 1 A280 Factor 2 A260 Factor 3 Pathlength 4 320nm background correction on off 5 Protein type 6 Dilution Factor 7 Unit 8 Sample Name Press OK to next step or Save to store the parameters Select between different sample...

Страница 27: ...ample before measurement 4 READ for measurement 5 PRN for result printing Note Press ZERO key after new setting this will be used for all subsequent samples SET Parameter setting Input 1 A280 Factor 2 A260 Factor 3 Pathlength 4 320nm background correction on off 5 Protein type 6 Dilution Factor 7 Unit 8 Sample Name Press OK to next step or Save to store the parameters Select between different samp...

Страница 28: ...return to previous screen Custom dye type selection For selected custom dye type Input 1 PRO EXT COEFFICIENT 2 Dye Abs Max for maximum absorbance wavelength of the custom dye 3 Dye Ext Coefficient for dye dependent extinction coefficient 4 Dye Correction for dye dependent correction factor at 280 nm Press OK to next step or Save to store the parameters ...

Страница 29: ...e before measurement 4 READ for measurement 5 PRN for result printing Note Press ZERO key after new setting this will be used for all subsequent samples SET Parameter setting Input 1 Wavelength 2 Cell Pathlength 3 Unit 4 Sample name 5 Dilution Factor 6 Multiplier from 1000 or 1 000 000 Press OK to next step or Save to store the parameters Select between different sample number ...

Страница 30: ...4 READ for measurement 5 PRN for result printing Note Press ZERO key after new setting this will be used for all subsequent samples SET Parameter setting Input 1 Data mode from OD Absorbance or T Transmittance 2 Pathlength 3 Sample Name 4 Select number of wavelength from 1 to 6 5 Wavelength nm need to measure Press OK to next step or Save to store the parameters Select between different sample num...

Страница 31: ...D for measurement 5 PRN for result printing Note Press ZERO key after new setting this will be used for all subsequent samples SET Parameter setting Input 1 Data mode from OD Absorbance T Transmittance or E S Energy 2 Start wavelength 3 End wavelength 4 Y Axis maximum to be shown 5 Y Axis minimum to be shown 6 Sample name Press OK to next step or Save to store the parameters 540nm 200 ...

Страница 32: ...w setting this will be used for all subsequent samples Data Process 1 P V Peak Valley function display the peak and valley 2 Smooth The smooth function employs a data smoothing algorithm to smooth the noise Selection of this function causes the spectral graph screen to change 3 Calculation K or K option performs addition multiplication of constant K on the spectral plot 4 Area for area calculation...

Страница 33: ...nt 5 PRN for result printing Note Press ZERO key after new setting this will be used for all subsequent samples SET Parameter setting Input 1 Data mode from OD Absorbance T Transmittance or E S Energy 2 Scan wavelength nm 3 Scan time period 4 Pathlength of the cuvette 5 Y Axis maximum to be shown 6 Y Axis minimum to be shown 7 Sample name Press OK to next step or Save to store the parameters ...

Страница 34: ...Note Press ZERO key after new setting this will be used for all subsequent samples Data Process 5 Kinetics Recalculate the kinetics by adjust the time period 6 Smooth The smooth function employs a data smoothing algorithm to smooth the noise 7 Calculation K or K option performs addition multiplication of constant K on the spectral plot 8 Area for area calculation ...

Страница 35: ...ult printing Note Press ZERO key after new setting this will be used for all subsequent samples SET Parameter setting Input 1 Pathlength 2 Sample name 3 Select number of wavelength from 1 to 3 4 Wavelength nm need to measure 5 Unit 6 Calibration mode Standard concentration or input K factor 7 Number of standard Press NEXT for Standard concentration or K factor input OK to next step or Save to stor...

Страница 36: ...ack to previous screen Concentration measurement After setting calibration curve Press 1 ZERO for blank or reference sample before measurement 2 READ for measurement 3 PRN for result printing Note Press ZERO key after new setting this will be used for all subsequent samples Change to standard curve display Select between different sample number ...

Страница 37: ...DNAmaster Quick Manual Version 1 0 Page 21 File Data IN SD 1 Select between data stored in internal memory or SD card 2 Open Delete the method or data stored ...

Страница 38: ...1 Home screen Select the function show on startup 2 Memory Storage Select storage by SD card or internal memory 3 Peak Threshold Select the threshold for automatic peak picking function 4 Factory reset reset back to factory default setting Time Time and Date setting ...

Страница 39: ...nal service tools may required please contact your local distributor for further assistance User 1 User Name input 2 Company name input 3 Display contrast of the screen 4 Energy saving mode for screen turn off after 10 minutes without using 5 Start up tune 6 Tune volume 7 Key Tone on off when pressing ...

Страница 40: ...ampling window and then back to the detector on the instrument It is supplied with a cap which contains a small mirror and it ensures that the sample is measured at constant pathlength The pathlength is achieved by a double pass through the sample as controlled by the cap height DNAmaster can setup the pathlength shown in the cap Just need to pipette a small drop 0 5 10μl depends on cap s pathleng...

Страница 41: ...0 ng μl 0 2 mm Lid Factor 50 ng μl 0 1 mm Lid Factor 100 ng μl Total Detection Range ng μl dsDNA 50 6 425 13 850 63 4250 125 8500 6 8500 ssDNA 37 5 315 9 629 46 3145 93 6290 5 6290 RNA 40 5 340 10 680 50 3400 100 6800 5 6800 Oligo 33 4 281 8 561 41 2805 83 5610 4 5610 Corresponding Absorbance Value Limits for A 260 Abs 0 025 to 1 7 0 025 to 1 7 0 025 to 1 7 0 025 to 1 7 Required Sample Volume 6 10...

Страница 42: ... temperature control function TM function For analyzing DNA melting temperature at 260nm select the OD600 TM function in the main menu Press 1 SET for detail setting 2 LOAD for reload saved parameter 3 ZERO for blank or reference sample before measurement 4 READ for measurement 5 PRN for result printing Note Press ZERO key after new setting this will be used for all subsequent samples ...

Страница 43: ...then 2 min 5 Starting temperature minimum 20 6 End temperature maximum 95 7 Data resolution 8 Set EQUIL TIME for time lag after reaching start temperature before reading start 9 Set K factor i e GC content K Tm 69 3 x 2 44 Press OK to next step or Save to store the parameters SYSTEM The temperature control model can also used for sample incubation or temperature control In Setting menu input 1 Tem...

Страница 44: ...ustralasia Dynamica Pty Ltd Unit C3 Hallmarc Business Park Corner Centre and Westall Roads Clayton Victoria 3168 Australia P 61 3 8540 5988 F 61 3 8540 5981 Email info dynamica com au Web www dynamica com au Asia Dynamica Asia Limited Room 607 Yen Sheng Centre 64 Hoi Yuen Road Kwun Tong Kowloon Hong Kong P 852 3583 1581 F 852 3583 1580 Email info dynamica asia com Web www dynamica asia com ...

Отзывы: