M
T
Cable specifications – curly cable
DX BUS Curly Cable
Tail Length La
Coil Length
Tail Length Lb
Part/Order number
200 mm / 8 inch
300 mm / 1 ft
200 mm / 8 inch
GSM63051
500 mm / 1'8"
300 mm / 1 ft
200 mm / 8 inch
GSM63052
R
-B
E
N
D
La ± 20
Ø26 ± 2
300 ± 30 appr
ox.
40 coils
compr
e
ssed coil lengt
h
Lb ± 20
Ø7.5
13 x 16
40
Parameter
Value
Contact Resistance (per contact)
As per IEC 512-2, Positronic
3 m
Ω
nom, 7 m
Ω
max
Wire resistance @ 20°C
DXB+
DXB-
15.1 m
Ω
/m, 4.6 m
Ω
/ft
8.5 m
Ω
/m, 2.6 m
Ω
/ft
Curly Cable electrical length
2.3 m / 7'6" nom for the coiled section
Operating Current (DXB+/DXB-)
12 A nom, 20 A RMS max
Connector Latch Holding Force
40 N min
Cable Strain
100 N max (accidental, non-repetitive)
Spring Force - extension refers to the
coiled section
< 20N @ 2x extension (T > 10°C/50°F)
< 50N @ 2x extension (T > –10°C/14°F)
< 30N @ 3x extension (T > 10°C/50°F)
Minimum Cable Bend Radius
Flexing values for (T > –10°C/14°F)
20 mm / 0.8 inch - fixed installation
30 mm / 1.2 inch - occasional flexing
50 mm / 2 inch
- frequent flexing
Operating Temperature
(ambient, fixed installation)
–25°C to +50°C
–13°F to +122°F
Cable Temperature Rating
80°C / 176°F (internal operating temp)
Warnings:
1.
Do not extend the spring coils below 0°C. Do not extend the coils
farther than 2x compressed length below 10°C, or 3x compressed
length above 10°C. Avoid extension above 3x compressed length
because this can cause a permanent stretching of the coil.
2.
Make sure that no spring force is applied to the DX BUS connector;
fasten a strain relief or cable tie on or near to the coiled section of
the cable.