background image

 

 

19 

Dymax ACCU-CAL™ 150 Radiometer User Guide 

Содержание ACCU-CAL 150

Страница 1: ...Dymax ACCU CAL 150 User Guide UV Radiometer Instructions for Safe Use Setup and Operation Maintenance Ordering Spare Parts and Accessories ACCU CAL 150 Radiometer PN 40550 ...

Страница 2: ...ow for 100 in line inspection and increase throughput System designs enable stand alone configuration or integration into your existing assembly line Please note that most dispensing and curing system applications are unique Dymax does not warrant the fitness of the product for the intended application Any warranty applicable to the product its application and use is strictly limited to that conta...

Страница 3: ...ting Your Shipment 7 Parts Included in the ACCU CAL 150 Radiometer 7 Operation 8 Setup and Default Modes 9 Explanation of Settings 10 Maintenance 10 General Maintenance 10 Cleaning and Calibration 10 Battery Replacement 12 Diagnostics and Error Messages 13 Spare Parts and Accessories 13 Radiometer Communications Software Kit PN 40629 13 Specifications 16 Warranty 17 New Product Warranty 17 Calibra...

Страница 4: ...questions or would like to help us with your suggestions or recommendations please contact our Application Engineering or Customer Service Departments at 860 482 1010 Trained Dymax professionals are standing by to serve you Si vous rencontrez un problème avait n importe de questions ou si vous voudrez de nous aider avec tes suggestions ou conseillations s il vous plait contacte notre departements ...

Страница 5: ...urité Emploi lunettes ou un protéger de visage pour protection de ultraviolet pour protéger vous oeux Chemises à manche long ou manteau de labo sont recommander pour protéger les bras et utilisation de ultraviolette gants opague vais protéger les mains Sicherheitshinweise Dieses Gerät wurde so entwickelt dass es nur vollständig alle Komponenten korrekt miteinander verbunden in Übereinstimmung mit ...

Страница 6: ...ings as total energy and dosage Environmental Considerations Optical measurement instruments are sensitive to extremes in environmental conditions like high temperature humidity and contamination Protect the device and its detector s from high humidity high temperature direct sunlight and contamination Do not use the ACCU CAL 150 immediately after moving it from a cold to a warm environment Under ...

Страница 7: ...and submit a claim for the damaged parts Contact Dymax so that new parts can be shipped to you immediately Check that the parts included in your order match those listed below If parts are missing contact your local Dymax representative or Dymax Customer Support to resolve the problem Figure 1 ACCU CAL 150 Radiometer PN 40550 shown in carrying case included Parts Included in the ACCU CAL 150 Radio...

Страница 8: ...ff 3 Entering RUN mode A short press of the RUN EXIT Button clears the memory and puts the unit in RUN mode The Display Window shows RUNNING after quickly displaying the internal temperature of the unit Confirm that the unit shows RUNNING before initiating a reading 4 Place the radiometer on the belt or object with the Optic Window Figure 3 looking toward the UV source The Display Window and butto...

Страница 9: ...e current settings Default modes will appear with an asterisk before them The Display Window below shows all possible choices for setup of each mode Figure 4 ACCU CAL 150 Display Window Run Mode Figure 5 ACCU CAL 150 Display Window Setup Mode SETUP MODE GRAPH REFERENCE TOGGLE SMOOTH OFF ON UNITS J W mJ mW uJ uW DISPLAYS LOW MEDIUM HIGH SAVE EXIT To change selections use the buttons under the down ...

Страница 10: ...as measured at 2 048 samples per second ON Displays the peak irradiance as measured at 25 samples per second Units Select the unit values J W mJ mW µJ µW Displays Select the display intensity LOW MEDIUM or HIGH Save Soft Button indicators and EXIT Maintenance General Maintenance The ACCU CAL 150 was designed to operate with minimal maintenance Follow the schedule below to ensure top unit performan...

Страница 11: ... a possible cause of the measurement error STEP 2 BLOW Use compressed gas to remove any loose material from the surface of the optics This step is necessary because loose material especially silicates and other abrasive components can cause scratching of the optics during the remaining steps Compressed gasses we recommend in order of preference are Dry nitrogen Chemtronics Duster PN ES1017 ES1217 ...

Страница 12: ...ing the optical surface with a dry wipe or swab of any kind The absence of solvent greatly increases the chance for the optical surface to be abraded Dymax can provide cleaning advice assistance and repair or replacement in cases where these guidelines fail to remove contaminants Following these guidelines will help customers get long life and accurate readings from their radiometers Dymax recomme...

Страница 13: ...be read by pressing the select button When doing this the beeping tone stops and you can scroll through the data readings In addition if the internal temperature of the unit exceeds 75 C the unit beeps once then displays the internal temperature continuously The unit will not operate until the internal temperature drops below 75 C The maximum internal temperature is 80 C If the internal temperatur...

Страница 14: ...e FINISH and or NEXT buttons until installation is complete Data Screen Button and Display Functions The ACCU CAL 150 Radiometer can be connected directly from the USB PC Connector Figure 7 on the side of the Radiometer to a USB Port on your PC via the supplied USB Cable Figure 7 USB PC Connector The Radiometer must be connected and turned on before transferring data Get Data Click this button to ...

Страница 15: ...r are updated The time and date stamp are from the PC Clock when GET DATA was pressed User Notes Any information added in this block by the end user will be retained and attached to the date for further reference Avoid using commas the tab key or the enter key in this box Error Messages Error 37 Incorrect Serial Port selected Go to SET UP and select the correct Serial Port number Error 43 Appears ...

Страница 16: ...tures for short periods audible alarm indicates when temperature has exceeded tolerance Time Out Period 2 minute DISPLAY mode no key activity Battery Two user replaceable AAA alkaline cells Battery Life Approximately 20 hours with display on Dimensions D x H 4 60 x 0 50 11 7 cm x 1 27 cm Weight 10 1 ounces 289 grams Instrument Materials Aluminum stainless steel Carrying Case Material Cut polyureth...

Страница 17: ...ll not apply to any accessory equipment or part that does not affect the proper operation of the device 2 Upon inspection the goods show evidence of becoming defective or inoperable due to abuse mishandling misuse accident alteration negligence improper installation lack of routine maintenance or other causes beyond our control 3 The goods have been repaired altered or modified by anyone other tha...

Страница 18: ...ill warranty calibration and or repair services just performed for 90 days This Calibration and Repair Warranty does not apply to nor cover repairs that may otherwise occur to the instrument Such repairs may be covered under the New Product Warranty based on the age of the instrument ...

Страница 19: ...19 Dymax ACCU CAL 150 Radiometer User Guide ...

Страница 20: ...ng equipment trial rental and leasing programs to assist in such testing and evaluations Data sheets are available for valve controllers or pressure pots upon request PN40682 MAN001 9 10 2012 Dymax Corporation 1 860 482 1010 info dymax com www dymax com Dymax Oligomers Coatings 1 860 626 7006 info_oc dymax com www dymax oc com Dymax Asia H K Limited 852 2460 7038 dymaxasia dymax com www dymax com ...

Отзывы: