Dymax BlueWave® LED DX-1000 VisiCure® User Guide
Table 4. Troubleshooting Chart for
BlueWave LED DX-1000
VisiCure
Problem
Possible Cause
Corrective Action
BlueWave LED DX-1000
VisiCure
does not power up
Power Cord not plugged
in or damaged
Check Power Connection and condition.
Onboard Fuse blown
Check Onboard Fuse.
No electrical power at
Receptacle
Test Receptacle for power.
BlueWave LED DX-1000
VisiCure
powers up but LED
does not produce light
LED intensity adjustment set
to minimum
Increase LED intensity setting.
Lens Cover is still in place
Remove Lens Cover. Clean Output Window if
necessary.
Interface Cable connections
loose or damaged
Check connections and condition of Interface
Cable.
BlueWave LED DX-1000
VisiCure
is operating
normally and LED suddenly
stops producing light
Over-temperature shutdown
was triggered
Check that the flow of cooling air into the
Controller or Irradiator Head unit is not restricted.
Check that the Filter is clean. If the LED does not
illuminate after restarting, contact Dymax
Applications Engineering.
LED light source provides
only low-intensity light
LED intensity adjustment set
to minimum
Increase LED intensity setting.
Contaminated/dirty Lens optics
Clean the surface of the Lens.
Footswitch does not
function
Footswitch not connected
Connect Footswitch.
Footswitch defective
Activate unit using the Front Control Panel.
Replace the Footswitch if the unit operates from
the Front Control Panel.
Unit appears to run hot
Cooling Fan Filter is dirty or
blocked
Replace Cooling Fan Filters. Remove items that are
blocking Filter and Air Inlet.
Insufficient clearance around
the Controller
Ensure 1" [2.54 cm] inch of clearance is provided
around the Controller and at least 6" [15.24 cm]
inches of clearance are provided around the
Cooling Fan Inlets.
Fan not operating
Ensure Fan is operating. Contact Dymax
Applications Engineering.
Controller is contaminated with
dust or debris
Contact Dymax Applications Engineering.