19
Dymax BlueWave® 75 User Guide
Troubleshooting
WARNING!
Only qualified maintenance personnel should attempt the following procedures.
AVERTISSEMENT!
Seulement personnel d’entretien diplomé devrais essayer les procedures suivant.
Table 1. Troubleshooting Chart for
BlueWave 75
Problem
Possible Cause
Testing
Corrective Action
Bulb will not
ignite
or
Change Bulb
Light is on
Improper connections
Visually inspect all input/output
connections (i.e. Power Cord,
Bulb).
Secure all connections.
Bulb is beyond useful
life of 2,000 hours
Replace with a new Bulb/Reflector
Assembly. Reset Bulb Hour Meter
and re-test unit.
Replace Bulb/Reflector Assembly if
required. Typical Bulb life is 2,000
hours.
Main line Fuse blown
(nothing in unit
operates)
Remove Fuse from Power
Receptacle and check with an
Ohmmeter.
Replace Fuse, if defective.
Low output
intensity
or
System fails
to cure
adhesive in
allotted time
Bulb beyond useful life
Use a Radiometer to measure
output intensity (Dymax ACCU-
CAL™ 50 or equivalent)
.
Replace Bulb/Reflector Assembly if
beyond useful life.
Transmission loss in
Lightguide too great
Compare Lightguide output against
new Lightguide (or use the Dymax
Lightguide Simulator) to determine
transmission loss.
Inspect and clean guide ends, and
eliminate tight bends.
Replace Lightguide.
Contaminants on
Lightguide
Visually examine ends of
Lightguide for contaminants.
Clean with isopropyl alcohol or
equivalent. Heavy deposits on
liquid Lightguides may be carefully
removed using a razor blade.
Replace the Lightguide if it cannot
be cleaned.
Bulb/Reflector
Assembly not installed
properly
Visually check to make sure the
Bulb/Reflector Assembly is seated
flush in the Bulb Mount Assembly
(any error in installation could
cause a low output)
Properly install Bulb/Reflector
Assembly, see page 17.