INSTALLATION

1.  Select a location that is free from excessive vibration, corrosive atmosphere and 

    where the ambient temperature is within the limits for these switches.

2.  Mount standard switches with the diaphragm in a vertical plane and with switch 

    lettering and nameplate in an upright position. Some switches are position 

    sensitive and may not reset properly unless they are mounted with the diaphragm 

    vertical. (Special units can be furnished for other than vertical mounting 

    arrangements if required.)

3.  Connect switch to source of pressure, vacuum or differential pressure. Metal tubing 

    with 1/4˝ O.D. is recommended, but any tubing which will not restrict the air flow 

    can be used. Connect to the two 1/8˝ female NPT pressure ports as noted below:

      •  Differential pressures - connect pipes or tubes from source of greater pressure 

        to high pressure port marked HI-PR and from source of lower pressure to low 

        pressure port marked LO-PR.

      • Pressure only (above atmospheric) - connect tube from source of pressure to 

        high pressure port. The low pressure port is left open to atmosphere.

      • Vacuum only (below atmospheric pressure) - connect tube from source of 

        vacuum to low pressure port. The high pressure port is left open to atmosphere.

4.  Electrical connections to the standard single pole, double throw snap switch are 

    provided by means of screw terminals marked “common”, “norm open”, and “norm 

    closed”. The normally open contacts close and the normally closed contacts open 

    when pressure increases beyond the set point.

5.  Switch loads should not exceed the maximum specified current rating of 15 amps 

    resistive. Switch capabilities decrease with high load inductance or rapid cycle 

    rates. Whenever an application involves either of these factors, the user may find 

    it desirable to limit the switched current to 10 amps or less in the interest of 

    prolonging switch life.

OPERATION

Pressure acting on the power diaphragm rotates the amplifying lever, which in turn 

extends the range spring and rotates the snap switch input lever. When the set point 

is reached, the snap switcch is actuated and the electrical contacts make or break.

ADJUSTMENT

To change the set point, proceed as follows:

  A.  Remove the snap-on cover from the conduit enclosure by loosening its retaining 

      screw and pulling firmly at its bottom end. Turn the slotted adjustment screw at 

      the top of range spring housing clockwise to raise the set point pressure and 

      counter-clockwise to lower the set point.

  B.  The recommended procedure for calibrating or checking calibration is to use a 

      “T” assembly with three rubber tubing leads, all as short as possible and the 

      entire assembly offering minimum flow restriction. Run one lead to the pressure 

      switch, another to the manometer of known accuracy and appropriate range, and 

      apply pressure through the third tube. Make final approach to the set point very 

      slowly. Note that manometer and pressure switch will have different response 

      times due to different internal volumes, lengths of tubing, fluid drainage etc. Be 

      certain the switch is checked in the position it will assume in use, i.e. with 

      diaphragm in a vertical plane and switch lettering and nameplate in an upright 

   position.

  C.  For highly critical applications it is a good idea to check the set point adjustment 

      and reset it as necessary once or twice in the first few months of operation. This 

      will compensate for any change in initial tension which may occur in the spring 

      and diaphragm. For most applications this change will not be significant and no 

      resetting will be required.

MAINTENANCE

Moving parts of these switches are sealed in and are permanently tamper proof. The 

single adjustment is that of the set point. Care should be taken to keep the switch 

reasonably dry and free from dust or dirt. No lubrication or unusual precautions are 

required for normal use.

Power must be off while wiring connections are being made.

CAUTION

Do not exceed the specified  voltage  rating. Permanent  damage 

not covered by warranty may result.

CAUTION

Содержание 1900 Series

Страница 1: ...EMOVAL 1 19 32 40 48 1 8 FEMALE NPT HIGH PRESSURE CONNECTION 3 76 20 ø 3 1 2 88 90 1 8 FEMALE NPT LOW PRESSURE CONNECTION 2 ø 3 16 4 76 MOUNTING HOLES ON A 4 3 16 106 36 B C 60 TYP 1 5 16 33 32 TYP 41 64 16 27 The Dwyer engineered force motion amplifier increases the leverage of diaphragm movement and results in a switch with excellent sensitivity and repeatability Series 1910 pressure switch All ...

Страница 2: ...terest of prolonging switch life OPERATION Pressure acting on the power diaphragm rotates the amplifying lever which in turn extends the range spring and rotates the snap switch input lever When the set point is reached the snap switcch is actuated and the electrical contacts make or break ADJUSTMENT To change the set point proceed as follows A Remove the snap on cover from the conduit enclosure b...

Страница 3: ...II sobretensión transitoria Conexiones a proceso 1 8 NPT hembra Orientación de montaje Diafragma en posición vertical Consulte con la fábrica para las orientaciones otra posición Ajuste del punto de ajuste Tipo de tornillo interior de la caja del conducto El grado de contaminación 2 Peso 1 lb 4 5 oz 581 g Aprobaciones de las agencias CE UL CSA y FM Opcional EXPL caja prueba de explosions no posee ...

Страница 4: ...CION La presión actúa sobre el diafragma girando la palanea amplificadora Esta a su vez extiende el resorte antagónico de calibrasión el que está mecánicamente vinculado a la palanca actuadora del interruptor inversor Al alcanzarse el valor preajustado la condición de los contactos se invierte es decir el NA pasa cerrado y el NC pasa abierto AJUSTE Para cambiar el ajuste proceda de la siguiente ma...

Отзывы: