8
1
INSTALLATION
7. Set-up alternatives for
Regeneration Unit
•
Set-up of Regeneration Unit RU 24 next
to the X-ray developer:
Where sufficient space is available, this
set-up is preferred.
•
Set-up of Regeneration Unit RU 24 in a
cupboard:
The operator must be instructed to check
the state of the chemicals regularly.
Set-up of Regeneration Unit
RU 24 above the level of the X-ray
developer is not permissible. This
set-up position makes it possible
that chemicals leak into the X-ray
developer baths.
Please ensure that all connecting hoses from
the Regeneration Unit RU 24 to the X-ray
developers are laid without any twists or
bends. The distance between Regeneration
Unit and X-ray developer should not exceed
2m.
When setting-up the X-ray developer,
ensure good access to it and to the
collecting tanks.