24
2160100227L02 1910V008
Assembly
DC emitter, 6 mA
Tube length 30 cm
Without X-ray field
limitation
X-ray field limitation
2x3
X-ray field limitation
3x4
60 kV
mGycm
2
60 kV
mGycm
2
60 kV
mGycm
2
Incisors
010 s
91
010 s
19
010 s
38
Premolars
013 s
118
013 s
25
013 s
50
Molars
016 s
148
016 s
31
016 s
62
Bite wing
016 s
148
016 s
31
016 s
62
i
Check and adjust the specific X-ray unit in accordance with the standard values
8.5 Acceptance tests
The required tests (eg acceptance tests) must be carried out in accordance with local rules and reg-
ulations
i
Find out which tests are required
i
Carry out testing in accordance with local rules and regulations
Acceptance test
The Intra / Extra Digital test body is required for acceptance tests with the image plate and
sensor as receivers, and possibly also the corresponding test body holder
i
Before the unit is started up and used for the first time, the acceptance test of the X-ray system
must be carried out in accordance with national regulations
The consistency tests, which must be carried out at regular intervals by the dental professional, are
based on the results of the acceptance test
Electrical safety checks
i
Carry out the electrical safety check according to the national law (e g in accordance with IEC
62353)
i
Document the results
i
Carry out and document the instruction and handover for the unit
A sample handover report is included in the attachment