![Durr Dental VistaScan Nano Easy Скачать руководство пользователя страница 20](http://html.mh-extra.com/html/durr-dental/vistascan-nano-easy/vistascan-nano-easy_installation-and-operating-instructions-manual_2542939020.webp)
18
2160100227L02 1910V008
Assembly
i
Remove the screw covers and screws from
the bottom of the device
i
Remove the cover from the rear of the device
7.2 Electrical connections
Safety when making electrical connections
i
The device must only be connected to a cor-
rectly installed power outlet
i
Do not place non-fixed multi-socket units on
the floor Follow the requirements in sec-
tion 16 of IEC 60601-1 (EN 60601-1)
i
Do not operate any other systems using the
same multiple socket
i
Make sure that none of the electrical cables
leading to the unit are under any mechanical
tension
i
Before initial start-up check that the mains
supply voltage and the voltage stated on the
type plate match (see also "4 Technical
data")
Connecting the unit to the mains supply
The unit has no main power switch For
this reason it is important that the unit is
be set up in such a way that the plug can
be easily accessed and unplugged if re-
quired
Requirements:
– Properly installed power outlet close to the
unit (observe the max mains cable length)
– Easily accessible power outlet
– Mains voltage must match the information
shown on the type plate of the power supply
unit
Only the supplied power supply unit may
be used
i
Attach the matching country-specific adapter
to the power supply unit