Important information
2.4 General safety information
i
When operating this device always observe all
guidelines, laws, and other rules and regula-
tions that are applicable at the site of opera-
tion�
i
Prior to each use, check condition of the de-
vice and make sure it is in perfect working or-
der�
i
Do not convert or modify the units�
i
Observe the Installation and Operating In-
structions�
i
Make the Installation and Operating Instruc-
tions available to the person operating the de-
vice at all times�
2.5 Combining devices safely
Take care when connecting units together or to
parts of other systems as there is always an ele-
ment of risk (e�g� due to leakage currents)�
i
Only connect units when there can be no
question of danger to operator or to patient�
i
Only connect units when it is safe to do so
and there is no risk of damage or harm to the
surroundings�
i
If it is not 100% clear from the unit data sheet
that such connections can be safely made or
if you are in any doubt, always get a suitably
qualified person (e�g� the manufacturer) to ver-
ify that the setup is safe�
Where applicable, the requirements for medical
products have been taken into account in the
development and construction of the device� As
a result, this device is suitable for installation
within medical supply equipment�
i
Where this device is installed within other
medical supply equipment, the requirements
set out in Directive 93/42 EEC and the rele-
vant standards must be complied with�
A copy of the system manufacturer‘s
declaration in accordance with Article
12 of Directive 93/42/EEC can be
found in our download section at
www�duerrdental�com (document no�
9000-461-264)�
2 Safety
Dürr Dental has designed and constructed this
device so that when used properly and for the
intended purpose there is no danger to people
or property� Nevertheless, residual risks can re-
main� You should therefore observe the follow-
ing notes�
2.1 Intended purpose
The suction unit provides the dental treatment
unit with vacuum and volume flow�
2.2 Intended use
Working in combination with the suction unit
with treatment unit, suction handpiece and can-
nula, the media used in dental treatment (e�g�
water, saliva, dentine and amalgam) are re-
moved by suction for disposal�
This unit is technically suitable for the aspiration
of nitrous oxide (laughing gas)� However, when
assembling a system for aspiration of nitrous
oxide, it is important to ensure that the other
components in the system are also suitable for
this purpose� Those responsible for setting up
the system must assess this and approve and
release the system for the aspiration of nitrous
oxide�
Operation with nitrous oxide is only per-
mitted if the exhaust air is transported
from the unit to the outside of the build-
ing�
2.3 Improper use
Any other usage or usage beyond this scope is
deemed to be improper� The manufacturer ac-
cepts no liability for damages resulting from this�
In these cases the user/operator will bear the
sole risk�
i
Do not use this device to aspirate flammable
or explosive mixtures�
i
The unit must not be used as a vacuum
cleaner�
i
Do not use chemicals containing chlorine or
foaming chemicals�
i
Operation in operating theatres of explosive
areas is not permissible�
i
The following suction units must not be set up
in the patient environment (within a radius of
1�5 m): V 600, V 900 S and V 1200 S�