Combining devices safely
– The overall safety of the unit and its main per-
formance features are independent of the net-
work. The device is designed for operation
independent of a network. However, some of
the functions are not available in this case.
– Incorrect manual configuration can lead to sig-
nificant network problems. The expert knowl-
edge of a network administrator is required for
configuration.
– The data connection utilises part of the band-
width of the network. Interactions with other
medical devices cannot be completely ruled
out. Apply the IEC 80001
‑
1 standard for risk
assessment.
– The device is not suitable for direct connection
to the public internet.
Network configuration
Various options are available for network configu-
ration:
ü
Automatic configuration via DHCP (recom-
mended).
ü
Automatic configuration via Auto-IP for direct
connection of unit and computer.
ü
Manual configuration.
❯
Configure the network settings of the unit using
the software or, if available, the touch screen.
❯
Check the firewall and release the ports, if
applicable.
Network protocols and ports
Port
Purpose
Service
45123 UDP,
45124 UDP
Unit recognition and
configuration
1900 UDP
Service detection
SSDP /
UPnP
502 TCP
Unit data
514
1)
UDP
Event protocol data
Syslog
22 TCP
Diagnosis
Telnet,
SSH
123 UDP
Time
NTP
1)
The port can vary depending on the config-
uration.
10 Adjustment options
10.1 Adjustment of the switch-
on/cut off pressure
WARNING
Risk of explosion of the pressure ves-
sel
The pressure vessels used in the com-
pressors are designed to withstand
continuous pressure changes of 2 bar
and can be used continuously under
these pressure changes.
❯
For load changes > 2 bar (max. per-
missible: 3 bar), comply with the maxi-
mum load change cycles specified in
the operating instructions of the pres-
sure vessel.
The pressure adjustment is performed in standby
mode.
❯
Press the standby button
for at least 2 sec-
onds.
❯
Touch the service key
for at least 2 sec-
onds.
The blue LEDs in the operating panel flash.
They are touch-sensitive and can be adjusted
accordingly.
5
6
7
8
9
bar
The pressure adjustment is performed in
0.5 bar increments by touching the LED.
❯
Touch the first flashing LED with your finger
and swipe to the required
switch on pressure
.
❯
Touch the last flashing LED with your finger and
swipe to the required
cut off pressure
.
❯
Confirm with the service key
.
If no touch pulse is received for 30 sec-
onds, the system will automatically switch
to standby operation. The settings are not
saved.
Assembly
5286100072L02 1909V005
19
EN