23
Troubleshooting
LCD Display
Information
Reason / Meaning
Err is displayed in the
SYS area, 1 is displayed
in the DIA area
The leakage is too fast or the pulse signal is too
weak. Please check the cuff, re-fasten it around
your arm and try again.
Err is displayed in the high
SYS area, 2 is displayed
in the DIA area.
The blood pressure signal cannot be detected
due to too much noise. Please remove the noise
sources and measure again.
Err is displayed in the
SYS area, 3 is displayed
in the DIA area.
The test result is abnormal. Please measure again.
Err is displayed in the
SYS area, P is displayed
in the DIA area.
The inflation fails. Please check the cuff, re-fasten it
around your arm and try again.
HI is displayed in the DIA
area.
The inflation pressure is greater than 307±8 mmHg
( 41 kPa±1 kPa ). Please measure again.
This section includes a list of error messages and frequently asked questions for
problems you may encounter with your blood pressure monitor. If the monitor is
not operating as you think it should, check here before arranging for servicing.
Содержание BPM200
Страница 1: ...BPM200 Blood Pressure Monitor PLEASE READ ALL THE INSTRUCTIONS BEFORE USE ...
Страница 2: ...2 ...
Страница 4: ...4 ...
Страница 5: ...5 English Manual ...
Страница 32: ...32 Manuel d utilisation en français ...
Страница 34: ...34 Informations de sécurité ...
Страница 35: ...35 ...
Страница 55: ...55 Gebrauchsanleitung auf Deutsch ...
Страница 78: ...78 Manual De Instrucciones En Español ...
Страница 101: ...101 Manuale D istruzioni In Italiano ...
Страница 124: ...124 Instrukcja obsługi Polski ...
Страница 147: ......