- 22 -
12. Trouble Shooting
(1)
Possible error messages and trouble shooting
Error
Error description
Trouble shooting
door
Door open or door switch error
Close the door and check the
switch
ProG
Program setting value error
Check setting values.
Err01
Disconnected the sensor
Replace the sensor
Err02
Board or control IC is damaged
Inspection control PCB
Err03
Heating elements damaged
Replace heating elements
Err04
Short sensor or no heat
Inspection control PCB sensor part
and check heating element