D U N I W A Y S T O C K R O O M C O R P .
23 of 30
2. Units Of Measurement
To read the chosen units of measure (both gauges), Send “u” (ASCII value 117); the 960 returns
Torr
or
mBar
or
Pasc
3. Set Point #1
To read the setting and status of set point #1
Send “1” (ASCII value 49); the 960 returns information for set point #1 in the format:
ABeC, DEeF, R, XXX
where:
AB is the multiplier and C is the exponent for set point #1 high
DE is the multiplier and F is the exponent for set point #1 low
R is the set point relay status; 0=relay is not energized, 1=relay is energized
XXX is the gauge to which the set point #1 has been assigned: either CCG or CVT.
4. Set Point #2
To read the setting and status of set point #2
Send “2” (ASCII value 50); the 960 returns information for set point #2 in the same format as for
set point #1, above.
5. Model And Software Revision
To read software identification.
Send “v” (ASCII value 118); the 960 returns the model number of the instrument and the revision
number, as in the following example:
960,ver. 1.10x
6. Factory Diagnostic
A diagnostic command is included for factory use only, which produces a repeating stream of
data.. It is initiated by sending “c” and terminated by sending “e”, or cycling the unit input power.
For additional information, please call Duniway Stockroom Corp.
7. Serial RS-232 Cable for PC Computer
Figure 15: Serial RS-232 Cable for PC Computer