RFHU Operator’s Manual - Shelf Control
10
RFHU OPERATING INSTRUCTIONS SHELF CONTROL- Continued
The tools menu will also contain the “TEMP” option
which allows the user to switch between Celsius and
Fahrenheit display.
The default is Fahrenheit (F).
When modified all temperatures that appear on the
runtime screen (actual and preset), grid display or
the preset editor will be displayed is the selected
temperature mode.
To lock or unlock the modification of the preset
value, first enter the admin menu by selecting the
“Admin” and entering the “0654” pin code when
prompted..
Follow this by selecting the “Access” option and then
use the slider to either lock or unlock the preset
editing -
The temperature calibration option is also accessible
from the “Admin” menu to edit the offset value for
each zone on the unit. When this menu option is
selected the current grid display will be displayed
with each of the individual temperature zones.
Select a zone to edit the temperature calibration
offset for the selected zone.
The temperature offset can be modified. The user
will be allowed to edit the temperature calibration
offset value from -20 F to 20F.
Use the and to increase or decrease the tem-
perature calibration offset value.
When done select to save the offset value else
select the button to exit without saving.
The
Help
option is reserved for future use.
HELP