9

Fault

Possible cause

Solutions

There is no water coming 

out from the handpiece 

(automatic water supply 

mode).

There is air in the water pipe.

Turn the water control to the Max, reinsert the bottle.

If the problem still can't be solved, please contact with local dealer or manufacturer.

Manufacturer will provide circuit diagrams, component part lists, descriptions, calibration instructions to assist to service 

personnel in parts repair.

3.6.2 Notice 

 

[Note 1]  The water control knob can adjust the water volume according to the  symbol.

[Note 2] To clean the water pipe with the multi-function syringe of the dental unit ( see picture 7):

Multi-function syringe

Picture 7

a) Snip the water pipe at a distance of 10cm to 20cm from the water entrance.

b) Turn on the power switch, get through to the power.

c) Connect the multi-function syringe of the dental unit to the water pipe.

d) Screw off the scaling tip or pull out the handpiece.

e) Step on the foot switch.

f) Turn on the switch of the multi-function syringe, press the air or water into the water pipe to clean and eliminate the impurity.

Содержание S6 LED

Страница 1: ...S6 LED ULTRASONIC SCALER INSTRUCTION MANUAL...

Страница 2: ...llation and adjustment 5 3 Maintenance and sterilization 6 4 Precaution 10 5 After service 12 6 Symbol instruction 12 7 Environmental protection 13 8 Manufacturer s right 13 9 European authorized repr...

Страница 3: ...nce and structural composition Ultrasonic scaler S6 LED is composed of electrocircuit water way and ultrasonic transducer 1 2 3 Scope of application Ultrasonic scaler S6 LED is used for the dental cal...

Страница 4: ...itrous Oxide Equipment not suitable for being used in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide 1 3 17 The main unit use a power adapter to connet network power s...

Страница 5: ...3 1 4 2 Back of the main unit sketch map Picture 2 1 4 3 Instruction of using wrench Picture 3...

Страница 6: ...4 Picture 4 1 4 4 List of removable parts Picture 5 Picture 6...

Страница 7: ...ee picture 4 2 1 6 Turn on the power switch the power indicator lighted and the machine is ready for work Touch panel is applied to this machine power can be adjusted by directly touching the power in...

Страница 8: ...notice whether the outer of the handpiece is damaged during the treatment or sterilization don t smear any protective oil on the surface of handpiece d There are two waterproof o rings at the end of...

Страница 9: ...while touching teeth 3 5 Cleaning of tips endochuck torque wrench and endo wrench The scaling tip endochuck torque wrench and endo wrench can be cleaned by ultrasonic cleaner 3 6 Troubleshooting and n...

Страница 10: ...er There is impurity in the solenoid valve Contact with the local distributor or manufacturer The handpiece generates heat The amount of spouting water is too little Turn the water control switch to a...

Страница 11: ...2 Notice Note 1 The water control knob can adjust the water volume according to the symbol Note 2 To clean the water pipe with the multi function syringe of the dental unit see picture 7 Multi functi...

Страница 12: ...g out Please keep the water flow smoothly 4 1 9 Don t use impure water source and be sure not to use normal saline instead of pure water source 4 1 10 If use the water source without hydraulic pressur...

Страница 13: ...a room where the relative humidity is 10 93 atmospheric pressure is 70kPa to106kPa and the temperature is 20 C 55 C 4 3 4 Please turn off the power switch and pull out the power plug when the equipme...

Страница 14: ...Symbol instruction Trademark Ordinary equipment Alternating current Drip proof Date of manufacture Manufacturer Class II equipment Type B applied part Foot switch interface Used indoor only Adjustment...

Страница 15: ...the electromagnetic environment specified below The customer or the user of the models S6 LED should assure that it is used in such an environment Emissions test Compliance Electromagnetic environmen...

Страница 16: ...rruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec...

Страница 17: ...ansmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range b Interference may occur In the vicinity of equipment marked with the follow...

Страница 18: ...2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to...

Страница 19: ...ZMN WI 09 416 V1 2 20181102...

Отзывы: