Error! No text of specified style in document.
43
3.
After the device has been on for 1-4 hours, place the device in the packaging tray
onto its assigned receiver.
4.
From the
Setup
menu, choose
PD Setup
... and change the analysis module to BP
for the pressure channel in the Channel Input Setup screen.
5.
Close the
Setup
window and start an acquisition from
Acquisition
– Start
Acquistion.
6.
In the
Status
window, double-click on the blood pressure channel to open the Blood
Pressure Analysis Attributes window. Select the
Offsets
tab.
7.
Under “Implant Pressure Offset”, click the
Measure
button to see an instantaneous
reading of the offset from 0. Clicking this again will give you an updated instantaneous
offset. Click
Cancel
to close this window without readjusting the pressure trace.
8.
Record the offset value on the Lab Sheet. If desired, a hard copy of a trace can be
made by clicking on the print button in the graph page menu bar. This should be kept with
the lab data for the project as verification of initial accuracy.