Operator’s Manual

Model 

614B • 

Laboratory Centrifuge

Rev. C

P/N 7711001

Reliability in Laboratory Centrifuges Since 1932

Содержание 614B

Страница 1: ...Operator s Manual Model 614B Laboratory Centrifuge Rev C P N 7711001 Reliability in Laboratory Centrifuges Since 1932 ...

Страница 2: ...f both the operator and service personnel care should be taken when using this centrifuge if handling substances that are known to be toxic radioactive or contaminated with pathogenic microorganisms When Risk Group II materials are used as identified in the World Health Organization Laboratory Bio Safety Manual a Bio Seal should be employed The Bio Seal accessory for the model 614 tube holders is ...

Страница 3: ...Maximum capacity 90 mL 6 x 15 mL Overall Dimensions H x W x D 8 5 in x 11 in x 12 5 in Centrifuge Motor 1 30 HP AC Nominal Acceleration Time 45 seconds Protection Breaker 4 Amp re settable Timer mechanical 1 to 30 minutes accuracy 10 Current Requirement 1 0 Amps Voltage Requirement 115 10 Volts Frequency 60 Hz Weight 18 5 lbs Before using any cleaning or decontamination methods except those recomm...

Страница 4: ...s be properly grounded 2 Turn the latch counter clockwise and open the lid 3 Spin the rotor by hand check for free and level rotation 4 Close the lid Rotate the lid knob clockwise to its complete stop position 5 Turn the centrifuge on by turning the timer to 10 minutes 6 Listen to the centrifuge A smooth whirring sound should be heard After the centrifuge has passed this procedure it is ready for ...

Страница 5: ... the top of the cabinet by a latching knob and pawl system When the knob is rotated clockwise the pawl grips the underside of the cabinet opening and prevents the lid from opening A mechanical stop positions the pawl and prevents it from rotating completely When rotated to the stop position the pawl makes contact with a micro switch mounted underneath the cabinet top The lid safety switch prevents...

Страница 6: ... Model 614B uses a brushless A C motor It should not need servicig for the life of the centrifuge The electrical components are selected for high reliability and should not need service 5 Keep the Rotor Chamber Clean Every six months or whenever there is a tube breakage refer to the note in 3 it may be necessary to remove the rotor and clean the rotor chamber Follow the instructions on page 5 to r...