![Drucker Diagnostics DASH Flex Series Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/drucker-diagnostics/dash-flex-series/dash-flex-series_operators-manual_2533132017.webp)
Drucker Diagnostics
–
Customer Service: +1-814-692-7661
–
[email protected] | Page 13
G
ENERAL
S
PECIFICATIONS
The rotor and accessories are rated for the maximum rotation frequency shown in the table below.
HORIZON 12 Flex
HORIZON 24 Flex
Tube Capacity
12 tubes
–
3 to 10 mL
24 tubes
–
3 to 10 mL,
12 tubes
–
15 mL
6 tubes - 50 mL
Dimensions (H x W x D)
13 in x 15 in x 9 in
(33 cm x 38 cm x 23 cm)
15 in x 17 in x 9 in
(38 cm x 43 cm x 23 cm)
Weight
34 lbs. (15 kg)
39 lbs. (17 kg)
Sound Level
56 dB A
59 dB A
Environmental Range
16
–
32 degree C
16
–
32 degree C
Voltage
95 -253 VAC
95 -253 VAC
Frequency
50/60 Hz
50/60 Hz
Power Requirement
280 Watts
280 Watts
Centrifuge Motor
½ H.P. Brushless
½ H.P. Brushless
Max g-Force
2,000 xg
2,000 xg
Max Speed
3,700 RPM
3,400 RPM
Cycle Time
.5 to 99 minutes (+/- 2%)
.5 to 99 minutes (+/- 2%)
C
ALCULATING THE
G-F
ORCE
The I.F.U.s of tube manufacturers recommend cycles at a minimum G-Force, which can be calculated if you know
the RPM and the radius. Use the formula below or go to www.druckerdiagnostics.com/g-force-calculator/.
In Centimeters:
RCF or G-force = 0.00001118 x
Rotor Radius (cm) x (RPM)
2
In Inches:
RCF or G-force = 0.0000284 x
Rotor Radius (in) x (RPM)
2
HORIZON 12 Flex
HORIZON 24 Flex
Radius
5.25 in (13.3 cm)
6 in (15.3 cm)