Drucker Diagnostics DASH Flex Series Скачать руководство пользователя страница 16

Drucker Diagnostics 

 Customer Service: +1-814-692-7661 

 [email protected] | Page 12  

Cycle parameters cannot be changed 

o

 

If cycle selection is locked on one cycle, press the UNLOCK 
button for 5 seconds. Then, press the GEAR button and follow 
the instructions elsewhere in this manual. 

o

 

If different saved cycles can be selected but not modified, the 
centrifuge is in Preset Lock mode. Press the UNLOCK button for 
5 seconds until two beeps are heard, then again until the next 
two beeps. You should now be able to change cycle parameters. 

The  centrifuge  does  not  unlock  after  a 
run is completed

 

 

o

 

Wait until the rotor has come to a complete stop. If the lid knob 
still cannot be rotated, press the UNLOCK button and try again.  

o

 

If no LED light is on, the unit is 

not powered and the lid will not 
unlock by conventional means. 
Remove the latch label and use a pen 
to manually disengage the locking 
mechanism. Pull the mechanism 
towards the control panel and then 
unlatch and open the lid. 

 

o

 

If the unit is damaged, contact Customer Service for assistance.

 

The lid does not open 

 

o

 

Ensure that the lid knob is turned fully counterclockwise. 

o

 

If the knob cannot be turned counterclockwise, turn it fully 
clockwise, press UNLOCK, and turn counterclockwise. 

o

 

If the lid remains locked after this and will not unlock, the 
electronics may have been damaged. Contact customer service 
for assistance. 

Clicking noise during braking gets loud 

o

 

Make sure that the screw in the center of the rotor is tight. 

Lid does not stay up 

o

 

Tighten the center screw on the lid hinge. 

 

 

 

Содержание DASH Flex Series

Страница 1: ...HORIZON 12 Flex 24 Flex Operator s Manual...

Страница 2: ......

Страница 3: ...Menu 4 Free Mode 5 Create New Cycle 5 Display Cycle Count 5 Modifying a Preset 6 Deleting a Cycle 7 Cycle Lock 7 Preset Cycles 7 Operation 8 Balancing Loads 8 12 Tube Centrifuges 8 24 Tube Centrifuge...

Страница 4: ......

Страница 5: ...being left in the centrifuge longer than necessary patent pending o A traditional audible alert indicates the completion of the cycle The audible alert can be muted o Cool Flow design prevents overhe...

Страница 6: ...is operator manual may cause interference to radio communications Operation of this equipment in a residential area may cause interference in which case the user will be required to correct the interf...

Страница 7: ...ry preset for Chemistry tubes 2 Coag PPP This setting is factory preset for Coagulation or Platelet Poor Plasma PPP 3 Urine This setting is factory preset for urine tubes Start Begins running the cycl...

Страница 8: ...d the CYCLE button until you hear a double beep The cycle number should begin flashing Enter the Advanced Menu Press the GEAR button to enter the advanced menu FROM FREE MODE Access the Menu Press the...

Страница 9: ...IME display Change Brake Values Access the menu Press the down key next to the top screen until it shows Brake To turn brake on and off use the up and down buttons next to the TIME display Exit the me...

Страница 10: ...CF button Setting Time Press the UP and DOWN buttons next to the TIME display Change Brake Values While in the advanced menu navigate to Brake Use the UP and DOWN buttons next to the TIME screen to tu...

Страница 11: ...changes to the selected cycle parameters The Preset Lock allows selection of any saved cycle and prevents changing the parameters of saved cycles Enter Single Cycle Select desired cycle With lid open...

Страница 12: ...ng mechanism will engage for 60 seconds allowing entry into the rotor chamber To unlock after more than 60 seconds have elapsed press the UNLOCK button The lid will unlock for another 15 seconds o Tur...

Страница 13: ...must be loaded symmetrically with tubes 15 mL 125 mm tubes may be spun in the standard HORIZON 24 Flex buckets but must be loaded as follows When three buckets are being used each bucket must be loade...

Страница 14: ...or storing the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is a...

Страница 15: ...n slight safely dispose of them and replace them o If the rotor appears damaged contact Customer Service for further assistance Excessive noise or vibration when the centrifuge is running o Verify tha...

Страница 16: ...annot be rotated press the UNLOCK button and try again o If no LED light is on the unit is not powered and the lid will not unlock by conventional means Remove the latch label and use a pen to manuall...

Страница 17: ...l Range 16 32 degree C 16 32 degree C Voltage 95 253 VAC 95 253 VAC Frequency 50 60 Hz 50 60 Hz Power Requirement 280 Watts 280 Watts Centrifuge Motor H P Brushless H P Brushless Max g Force 2 000 xg...

Страница 18: ...cling of waste electrical and electronic equipment The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is r...

Отзывы: