HORIZON
12 Flex, 24 Flex
Operator’s Manual
Страница 1: ...HORIZON 12 Flex 24 Flex Operator s Manual...
Страница 2: ......
Страница 3: ...Menu 4 Free Mode 5 Create New Cycle 5 Display Cycle Count 5 Modifying a Preset 6 Deleting a Cycle 7 Cycle Lock 7 Preset Cycles 7 Operation 8 Balancing Loads 8 12 Tube Centrifuges 8 24 Tube Centrifuge...
Страница 4: ......
Страница 5: ...being left in the centrifuge longer than necessary patent pending o A traditional audible alert indicates the completion of the cycle The audible alert can be muted o Cool Flow design prevents overhe...
Страница 6: ...is operator manual may cause interference to radio communications Operation of this equipment in a residential area may cause interference in which case the user will be required to correct the interf...
Страница 7: ...ry preset for Chemistry tubes 2 Coag PPP This setting is factory preset for Coagulation or Platelet Poor Plasma PPP 3 Urine This setting is factory preset for urine tubes Start Begins running the cycl...
Страница 8: ...d the CYCLE button until you hear a double beep The cycle number should begin flashing Enter the Advanced Menu Press the GEAR button to enter the advanced menu FROM FREE MODE Access the Menu Press the...
Страница 9: ...IME display Change Brake Values Access the menu Press the down key next to the top screen until it shows Brake To turn brake on and off use the up and down buttons next to the TIME display Exit the me...
Страница 10: ...CF button Setting Time Press the UP and DOWN buttons next to the TIME display Change Brake Values While in the advanced menu navigate to Brake Use the UP and DOWN buttons next to the TIME screen to tu...
Страница 11: ...changes to the selected cycle parameters The Preset Lock allows selection of any saved cycle and prevents changing the parameters of saved cycles Enter Single Cycle Select desired cycle With lid open...
Страница 12: ...ng mechanism will engage for 60 seconds allowing entry into the rotor chamber To unlock after more than 60 seconds have elapsed press the UNLOCK button The lid will unlock for another 15 seconds o Tur...
Страница 13: ...must be loaded symmetrically with tubes 15 mL 125 mm tubes may be spun in the standard HORIZON 24 Flex buckets but must be loaded as follows When three buckets are being used each bucket must be loade...
Страница 14: ...or storing the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is a...
Страница 15: ...n slight safely dispose of them and replace them o If the rotor appears damaged contact Customer Service for further assistance Excessive noise or vibration when the centrifuge is running o Verify tha...
Страница 16: ...annot be rotated press the UNLOCK button and try again o If no LED light is on the unit is not powered and the lid will not unlock by conventional means Remove the latch label and use a pen to manuall...
Страница 17: ...l Range 16 32 degree C 16 32 degree C Voltage 95 253 VAC 95 253 VAC Frequency 50 60 Hz 50 60 Hz Power Requirement 280 Watts 280 Watts Centrifuge Motor H P Brushless H P Brushless Max g Force 2 000 xg...
Страница 18: ...cling of waste electrical and electronic equipment The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is r...