This device is intended to be operated by properly trained personnel 
who have carefully read the operating manual and are familiar with the 
function of the device.  

 

 

For the safety of both the operator and service personnel, care should 
be taken when using this centrifuge if handling substances that are 
known to be toxic, radioactive or contaminated with pathogenic 
microorganisms. When Risk Group II materials are used, (as identified in 
the World Health Organization “Laboratory Bio-Safety Manual”), a Bio-
Seal should be employed. The Bio-Seal accessory for the model is the 
non-aerosol shield cap with appropriate tube holders (contact Drucker 
Diagnostics to purchase). In the event that materials of a higher risk 
group are being used, more than one level of protection must be 
provided. 

 

 

The use of flammable or explosive materials as well as those materials 
which have a vigorous chemical reaction is prohibited. 

 

 

For your safety and the durability of your machine, never transport or 
store centrifuge with tube holders inside machine. 

 

 

Inspect centrifuge for cracks or physical damage to cabinet, lid, rotor, or 

tube holders. Damage may result in unsafe operation. Discontinue use 

until repairs have been performed. 

 

Users should also comply with the specimen receptable manufacturer’s 
specific instructions for use, in addition to any other protocols 
established by the testing organization 

 

 

Содержание 642B

Страница 1: ...642B Operator s Manual P N 03 0 0002 0078 Rev H...

Страница 2: ...Setup Location 4 Initial Setup Procedure 5 Operation 6 Spinning Balanced Loads 7 Rotor Removal And Installation 7 To remove the rotor 7 To install the rotor 7 Care And Preventative Maintenance 8 Clean...

Страница 3: ...o Seal accessory for the model is the non aerosol shield cap with appropriate tube holders contact Drucker Diagnostics to purchase In the event that materials of a higher risk group are being used mor...

Страница 4: ...through the transparent lid In the event that the lid is opened during a run the power to the motor is disconnected This general purpose laboratory centrifuge may also be used to spin approved contain...

Страница 5: ...at prevents the centrifuge from operating unless the lid is closed and latched Removable rotor for easy cleaning Brushless permanent split capacitor AC motor Clear lid for safe observation of samples...

Страница 6: ...he center No person or hazardous material shall be permitted in the clearance envelope during operation The operator time within the envelope shall be limited to the time necessary for loading unloadi...

Страница 7: ...ers inside the rotor and verify that they are seated properly 4 Close the lid Rotate the lid knob clockwise to its complete stop position 5 Turn the centrifuge on by turning the timer to 10 minutes 6...

Страница 8: ...ders Be sure to follow the rules for balanced loads page 7 4 Close the lid and turn the lid knob clockwise to its complete stop position 5 Turn on the machine by turning the timer to the desired run t...

Страница 9: ...nplug the centrifuge from the electrical outlet to eliminate the possibility of electrical shock or other injury 2 Remove the test tube holders 3 Remove the nut in the center of the rotor by turning i...

Страница 10: ...safety guidelines of your laboratory to properly clean up and or dispose of materials in the event that a substance known to be potentially toxic radioactive or contaminated with a pathogenic microorg...

Страница 11: ...in water or other cleaning solutions as this will cause damage and void the warranty 3 ONLY isopropyl alcohol soap and water or a 10 5500 PPM bleach solution should be used for cleaning and disinfect...

Страница 12: ...is running o Check to see that the load is balanced o Make sure that nothing has fallen into the rotor chamber o Make sure that the nut in the center of the rotor is tight o Have a technician test the...

Страница 13: ...e to a complete stop and then unlock for 60 seconds If additional unlock time is needed press the OPEN STOP button with the machine plugged in and the rotor stopped If the lid remains locked after thi...

Страница 14: ...the top of the cabinet by a latching knob and pawl system When the knob is rotated clockwise the pawl grips the underside of the cabinet opening and prevents the lid from opening A mechanical stop pos...

Страница 15: ...ce horizontal 02 005 1 0006 Motor Assembly 1 30 H P 115 V A C permanent split capacitor 7729009 Capacitor 5uF 250V A C 7751043 Circuit Breaker 4A 7724071 Hinge friction 7732018 Seal lid gasket 02 002...

Страница 16: ...and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment For more...

Отзывы: