6.

 

Remove Accessories Before Moving:

 All tube holders, samples, 

and caps must be removed from the rotor chamber before 
transporting or storing the centrifuge to prevent damage and 
injury. 

C

LEANING AND 

D

ISINFECTION

 

To prolong the life of the centrifuge cleaning and disinfection is 
recommended every six months, or whenever there is a spillage or tube 
breakage. Contaminants must be removed immediately, or corrosion and 
premature degradation of components can occur. 

1.

 

Unplug the centrifuge before cleaning. 

2.

 

Apply cleaning solutions with a towel or cloth. Do not submerge 
the centrifuge in water or other cleaning solutions as this will 
cause damage and void the warranty. 

3.

 

ONLY isopropyl alcohol, soap and water, or a 10% (5500 PPM) 
bleach solution should be used for cleaning and disinfection of 
the centrifuge and accessories. 

4.

 

All surfaces must be dried immediately after cleaning and 
disinfecting. 

5.

 

TBQ Germicidal products shall not be used, as they will cause 
damage to the centrifuge and void the warranty. 

6.

 

The use of fully/partially halogenated hydrocarbons, ketones, 
esters, ethers, benzyls, ethyl benzenes, and all other chemicals 
not prescribed by the manufacturer shall not be used as they 
may cause damage to the rotor chamber, rotor, tube holders, 
accessories and centrifuge exterior and void the warranty. 

7.

 

It may be necessary to remove the rotor and clean the rotor 
chamber. Follow the instructions on page 8 to remove and 
reinstall the rotor. 

 

 

Содержание 642B

Страница 1: ...642B Operator s Manual P N 03 0 0002 0078 Rev H...

Страница 2: ...Setup Location 4 Initial Setup Procedure 5 Operation 6 Spinning Balanced Loads 7 Rotor Removal And Installation 7 To remove the rotor 7 To install the rotor 7 Care And Preventative Maintenance 8 Clean...

Страница 3: ...o Seal accessory for the model is the non aerosol shield cap with appropriate tube holders contact Drucker Diagnostics to purchase In the event that materials of a higher risk group are being used mor...

Страница 4: ...through the transparent lid In the event that the lid is opened during a run the power to the motor is disconnected This general purpose laboratory centrifuge may also be used to spin approved contain...

Страница 5: ...at prevents the centrifuge from operating unless the lid is closed and latched Removable rotor for easy cleaning Brushless permanent split capacitor AC motor Clear lid for safe observation of samples...

Страница 6: ...he center No person or hazardous material shall be permitted in the clearance envelope during operation The operator time within the envelope shall be limited to the time necessary for loading unloadi...

Страница 7: ...ers inside the rotor and verify that they are seated properly 4 Close the lid Rotate the lid knob clockwise to its complete stop position 5 Turn the centrifuge on by turning the timer to 10 minutes 6...

Страница 8: ...ders Be sure to follow the rules for balanced loads page 7 4 Close the lid and turn the lid knob clockwise to its complete stop position 5 Turn on the machine by turning the timer to the desired run t...

Страница 9: ...nplug the centrifuge from the electrical outlet to eliminate the possibility of electrical shock or other injury 2 Remove the test tube holders 3 Remove the nut in the center of the rotor by turning i...

Страница 10: ...safety guidelines of your laboratory to properly clean up and or dispose of materials in the event that a substance known to be potentially toxic radioactive or contaminated with a pathogenic microorg...

Страница 11: ...in water or other cleaning solutions as this will cause damage and void the warranty 3 ONLY isopropyl alcohol soap and water or a 10 5500 PPM bleach solution should be used for cleaning and disinfect...

Страница 12: ...is running o Check to see that the load is balanced o Make sure that nothing has fallen into the rotor chamber o Make sure that the nut in the center of the rotor is tight o Have a technician test the...

Страница 13: ...e to a complete stop and then unlock for 60 seconds If additional unlock time is needed press the OPEN STOP button with the machine plugged in and the rotor stopped If the lid remains locked after thi...

Страница 14: ...the top of the cabinet by a latching knob and pawl system When the knob is rotated clockwise the pawl grips the underside of the cabinet opening and prevents the lid from opening A mechanical stop pos...

Страница 15: ...ce horizontal 02 005 1 0006 Motor Assembly 1 30 H P 115 V A C permanent split capacitor 7729009 Capacitor 5uF 250V A C 7751043 Circuit Breaker 4A 7724071 Hinge friction 7732018 Seal lid gasket 02 002...

Страница 16: ...and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment For more...

Отзывы: