Cleaning 

The following guidelines are suggested by Drive DeVilbiss Healthcare as being suitable infection 
control procedures. Further information is available upon request.  

 

Pump Unit  

 

General cleaning may be affected by using a cloth dampened with a mild detergent and water 
solution. This approach may be followed either by wiping with a sodium hypochlorite solution to a 
dilution of 1000 ppm or by using an alcoholic wipe.  

 

 

WARNING:   Do not use phenolic or biological based cleaning solutions 

Do not use any abrasive compounds or cleaning pads 
Do not submerge the pump in to water or any other liquids 
Ensure that mains power supply to the pump has been disconnected prior to 
cleaning. 

 

Mattress

 

 

General Cleaning  

 

During general use, the mattress and tube set may be cleaned using a neutral detergent solution. 
Using a single use wipe, clean the mattress cover and/or tube set with a solution of neutral 
detergent and hot water. Rinse thoroughly with clean water and a damp single use wipe. 

 

NOTE:  The top, bottom and all four sides of the cover and tube set, including under the zip flaps 

MUST be cleaned.  

 

 

The foam used in this product cannot be decontaminated, therefore, it is recommended 
that any contamination must result in the foam being replaced 

 

 

 

WARNING:   Do not use phenolic or biological based cleaning solutions 

 

Do not use any abrasive compounds or cleaning pads 

Ensure the cover and tube set are completely dried before refitting the cover.   

 

Disinfecting the mattress   

 

If the cover is heavily soiled or has been exposed to bodily fluids such as blood, it will require a 
thorough cleaning procedure:  

 

Wipe the cover using a single use wipe with Chlorine Solution (of up to 10,000ppm) and cold water. 
Rinse thoroughly with clean water and a damp single use wipe. Make sure the cover is completely 
dried before refitting or storing.   

 

Wipe the cells using a single use wipe with Chlorine Solution (1,000ppm) and cold water.  Rinse 
thoroughly with clean water and a damp single use wipe. Make sure the cellsare completely dried 
before refitting or storing.   
 
Frequent or prolonged exposure to higher concentration disinfectant solutions may affect the 
longevity of the mattressand affect its functionality.    

 

NOTE:  Some surface wrinkling may also occur 

 

 

The cells in the system cannot be fully laundered due to the internal properties being 
affected.   

 

 

Personal Protective Equipment should be worn during the entire cleaning process 

 

 

This system must not steam cleaned or used in a autoclave 

Содержание DeVilbiss Healthcare Hybrid-Power

Страница 1: ...Hybrid Power User Manual...

Страница 2: ...p step down alternating therapy to repair and maintain healthy skin cells as well as providing a comfortable and stable platform When the Hybrid Power system is in use with the pump it is suitable for...

Страница 3: ...tent Fire Warning 3 Safeguards 4 Introduction 6 Installation 8 Pump Functions 9 Storage 10 Damage Prevention 10 Moving and Handling 10 Cleaning 11 Maintenance 13 Troubleshooting 14 Technical Descripti...

Страница 4: ...ause damage to the mattress system Patient clothing bed sheets and other items may be combustible and could catch fire Failure to observe this warning could result in a severe fire property damage phy...

Страница 5: ...to a service centre for examination and repair 5 Keep the cord away from heated surfaces 6 Never block the air opening of this product or place it on a soft surface such as a bed or couch where their...

Страница 6: ...5 Hybrid Power Disposal of Electrical Electronic Equipment WEEE Product conforms to all relevant standards See Declaration of Conformity...

Страница 7: ...cells to be built on as well as providing a support structure during patient transfer The support foam also contains specially engineered castellation s in the side walls to allow the mattress to ben...

Страница 8: ...unit This pump has been specifically designed for the Hybrid application This compact and light weight pump is easy to transport and can be easily cleaned down to aid infection control Intended use Th...

Страница 9: ...floor 6 Join the connectors on the mattress feed pipes to the side of the pump NOTE Ensure the feed pipes on the mattress are not kinked or tucked under the mattress as this will restrict airflow 7 P...

Страница 10: ...d by pressing this button The pump will automatically revert back to alternating mode after 20 minutes Alternate One in two cycle this function will provide optimum pressure relief Lock Out After 5 mi...

Страница 11: ...tra care not to puncture the cover as this will allow fluid ingress and ultimately contaminate the foam core To prevent the cover from being punctured and or damaged x Do not place sharp objects such...

Страница 12: ...ct cannot be decontaminated therefore it is recommended that any contamination must result in the foam being replaced WARNING Do not use phenolic or biological based cleaning solutions Do not use any...

Страница 13: ...ould be followed by a cold rinse and extraction Drying Mattress inners and tube sets should be wiped with a clean and dry cloth before reinserting Mattress covers can be tumble dried on a low heat set...

Страница 14: ...be checked regularly to ensure they remain fit for purpose clinically effective and pose no risk of infection to either the patient or the carer Drive DeVilbiss Healthcare Ltd recommend that a thorou...

Страница 15: ...ls are secure x Assess the cells for any punctures which maybe causing a leak Patient is bottoming out x Pressure setting may be inadequate adjust comfort range No air is being produced from some of t...

Страница 16: ...ge x Continuous operation Mattress Specification Dimensions 1980 x 880 x160mm Weight 16Kg Max Patient Weight 254kg with a pump Top Cover Material Xtreme fabric PU coated fabric substrate Cell Material...

Страница 17: ...our products and our assurance to the customer In the event of a defect covered by this warranty we will at our option supply parts or replace the device This warranty does not cover device failure du...

Страница 18: ...ve our customers the freedom to live independently This encompasses their daily home life and provides them with the opportunity to enjoy an outing with family and friends Our goal is to develop a ran...

Отзывы: