5. Pulse Rate Control
Pulse rate is adjustable from 2Hz to 150 Hz . Its value is visible on the LCD.
This buttons determine how many electrical impulses are applied through
the skin each second. By pressing these controls, the number of current
impulses per second (Hz) for both channels can be continually adjusted.
Unless otherwise instructed, turn the pulse rate control to the 70-120 Hz
range.
A. Increase
When pressing this button, the pulse rate will increase gradually.
B.Decrease
When pressing this button, the pulse rate will decrease gradually.
6. Pulse Width Control
Pulse Width is adjustable from 30 uS to 260 uS. Its value is visible on the
LCD. These buttons adjusts the length of time each electrical signal is
applied through the skin, which controls the strength and sensation of the
stimulation. If no
instructions regarding the pulse width are given in
therapy, set the control to the suggested 70-120 uS setting.
3. Lead Connector
Connection of the electrodes is made with two-lead connector. The device must
be switched off before connecting the cables. Both intensity controls must be at
the Off position. Electrodes must be pressed firmly on the skin.
4. Mode Control
Expose the controls by sliding front cover down from top of unit. This button
controlling 5 modes: B for burst stimulation, N for constant stimulation, and M for
pulse width modulation stimulation, S1 for intensity and pulse rate modulation,
S2 for intensity and pulse width modulation. Press the button until engaged in
mode desired.