Elite Mattress with PUMP1000
12
Technical Description
Pump
Specification
Power Supply
220-240V 50-60Hz 0.07A 16W
Fuse Rating
T1A / 250v
Cycle Time
10 minutes
Dimensions
291 x 200 x117 (mm)
Weight
2.8kg
Modes
Alternating, Static & Cushion
Air Output
8 litres per minute
Pressure Range
15mmHg
–
40mmHg±10%
Environment
Temperature
Operation
10°C to 40°C (50°F to 104°F)
Storage
-15°C to 50°C (5°F to 122°F)
Shipping
-15°C to 70°C (5°F to 158°F)
Humidity
Operation
20% to 90% (non-condensing)
Storage:
10% to 90% (non-condensing)
Shipping:
10% to 90% (non-condensing)
•
Not suitable for use in the presence of a flammable anaesthetic mixture (No AP or APG protection)
Noise Level
30 dB
Classification
o
Class I medical device
o
Type BF
Electrical classification
Class II
Ingress Protection
•
IP21 (Protection against fingers or other object not greater than 80mm
in length and 12mm in diameter) & (Vertically dripping water shall have
no harmful effect when the enclosure is tilted at an angle of 15° from its
normal position)
Applied Part
•
Elite Mattress & Elite Cushion
Usage
•
Continuous operation
Mattress
Specification
Dimensions
1880 x 890 x175mm
Weight
6Kg
Max Patient Weight
190kg
Top Cover Material
Xtreme fabric (PU coated fabric substrate)
Cell Material
Nylon / PVC
Cell Configuration
16 single cell formation
Fire Classification
BS5852 / BS7175 composite
–
(Risk assessment to be carried out)
Risk Factor
Very High Risk with pump
Electromagnetic Interference
•
This device has been tested to withstand electromagnetic interference (EMI) from
commonly found electrical equipment.
•
EMI can affect some types of electronic medical devices (such as pacemakers, defibrillators,
ECG monitors and infusion pumps). If in any doubt before use, a risk assessment should be
carried out before using this or any other electrical device around sensitive devices.
•
Electrical equipment may be susceptible to EMI from sources such as mobile phones, walkie-
talkies, TV broadcast and emergen
cy services’ radio. A risk assessment should also be carried
out before using electronic medical devices in areas close to these sources.
•
The addition of components or accessories may affect the EMI susceptibility of the device.
Do not fit accessories other than Drive DeVilbiss Healthcare authorised accessories.