Elite Mattress with PUMP1000 

 

Cleaning 

The following guidelines are suggested by Drive DeVilbiss Healthcare as being suitable infection 
control procedures. Further information is available upon request.  

 

Pump Unit  

 

General cleaning may be affected by using a cloth dampened with a mild detergent and water 
solution. This approach may be followed either by wiping with a sodium hypochlorite solution to a 
dilution of 1000 ppm or by using an alcoholic wipe.  

 

 

WARNING:   Do not use phenolic or biological based cleaning solutions 

Do not use any abrasive compounds or cleaning pads 
Do not submerge the pump in to water or any other liquids 
Ensure that mains power supply to the pump has been disconnected prior to 
cleaning. 

 

Mattress

 

 

General Cleaning  

 

During general use, the mattress and tube set may be cleaned using a neutral detergent solution. 
Using a single use wipe, clean the mattress cover and/or tube set with a solution of neutral 
detergent and hot water. Rinse thoroughly with clean water and a damp single use wipe. 

 

NOTE:  The top, bottom and all four sides of the cover and tube set, including under the zip flaps 

MUST be cleaned.  

 

 

The foam used in this product cannot be decontaminated, therefore, it is recommended 
that any contamination must result in the foam being replaced 

 

 

 

WARNING:   Do not use phenolic or biological based cleaning solutions 

 

Do not use any abrasive compounds or cleaning pads 

Ensure the cover and tube set are completely dried before refitting the cover.   

 

Disinfecting the mattress   

 

If the cover is heavily soiled or has been exposed to bodily fluids such as blood, it will require a 
thorough cleaning procedure:  

 

Wipe the cover using a single use wipe with Chlorine Solution (of up to 10,000ppm) and cold water. 
Rinse thoroughly with clean water and a damp single use wipe. Make sure the cover is completely 
dried before refitting or storing.   

 

Wipe the cells using a single use wipe with Chlorine Solution (1,000ppm) and cold water.  Rinse 
thoroughly with clean water and a damp single use wipe. Make sure the cells are completely dried 
before refitting or storing.   
 
Frequent or prolonged exposure to higher concentration disinfectant solutions may affect the 
longevity of the mattress and affect its functionality.    

 

NOTE:  Some surface wrinkling may also occur 

 

 

 

 

Содержание Elite

Страница 1: ...Elite with Digital Pump User Manual ...

Страница 2: ...both the prevention and treatment of pressure ulcers in any healthcare environment With a fully encased layer of visco elastic foam beneath the cells for added support the Elite mattress combines a comfortable support surface and alternating therapy suitable for patients at Very High Risk of developing pressure ulcers ...

Страница 3: ...Content Fire Warning 3 Safeguards 4 Intended Use 5 Contradictions 5 Installation 6 Pump Functions 7 Mattress Functions 8 Storage 8 Damage Prevention 8 Cleaning 9 Maintenance 11 Technical Description 12 Troubleshooting 13 Warranty 14 ...

Страница 4: ...rning could result in a severe fire property damage physical injury or death DO NOT use candles on or around the mattress system DO NOT use electric blankets with this mattress system DO NOT use in the presence of flammable anaesthetics explosions could occur DO keep hot equipment off and away from the mattress surface E g Hair dryer curling tong etc DO keep matches and lighters away from children...

Страница 5: ...dropped or damaged or dropped in water return the product to a service centre for examination and repair 5 Keep the cord away from heated surfaces 6 Never block the air opening of this product or place it on a soft surface such as a bed or couch where their openings may be blocked Keep the air opening free of lint hair and other similar particles 7 Never drop or insert any object into any opening ...

Страница 6: ...tended Use The Elite Replacement System is intended to reduce the incidence of pressure ulcers while optimising patient comfort It may also be appropriate for the following Individual home care setting and long term care facilities Pain management as prescribed by a healthcare professional Contraindications Users with the following condition could experience contraindications if they use this devi...

Страница 7: ...o secure the pump against the foot board to reduce vibration Avoid placing the pump on the floor 6 Join the connectors on the mattress feed pipes to the side of the pump NOTE Ensure the feed pipes on the mattress are not kinked or tucked under the mattress as this will restrict airflow 7 Plug the pump into the mains electrical outlet NOTE Any extra power cord should be neatly arranged and off the ...

Страница 8: ...ic All cells in the mattress will be inflated by pressing this button The pump will automatically revert back to alternating mode after 20 minutes Alternate One in two cycle this function will provide optimum pressure relief Patient Weight Settings 10 comfort levels are available for a variety of patient weights Use the to adjust the weight settings accordingly Low Pressure Indicator When low pres...

Страница 9: ...and stored in a holdall or bag To prevent mattresses becoming contaminated and or damaged during storage Mattresses should be stored within a polythene cover or bag Do not store other objects such as side rails or bed ends on top of mattresses Allow adequate space for the stored mattresses to be manoeuvred Damage Prevention Take extra care not to puncture the cover and cells as this will allow flu...

Страница 10: ...ottom and all four sides of the cover and tube set including under the zip flaps MUST be cleaned The foam used in this product cannot be decontaminated therefore it is recommended that any contamination must result in the foam being replaced WARNING Do not use phenolic or biological based cleaning solutions Do not use any abrasive compounds or cleaning pads Ensure the cover and tube set are comple...

Страница 11: ...attress covers can be tumble dried on a low heat setting for 90 minutes Drying temperature must not exceed 40 C Exceeding this temperature can cause significant damage to the mattress cover NOTE Covers must be completely dried before refitting to the mattress inners Biocides This product contains a biocide anti fungal agent to control microbial deterioration The active ingredient is silver phospha...

Страница 12: ...ses should be checked regularly to ensure they remain fit for purpose clinically effective and pose no risk of infection to either the patient or the carer Drive DeVilbiss Healthcare Ltd recommend that a thorough inspection of both the interior foam base and cells and exterior cover of the mattress is carried out on a weekly basis or each time a new patient is placed on the mattress Visual checks ...

Страница 13: ...Elite Cushion Usage Continuous operation Mattress Specification Dimensions 1880 x 890 x175mm Weight 6Kg Max Patient Weight 190kg Top Cover Material Xtreme fabric PU coated fabric substrate Cell Material Nylon PVC Cell Configuration 16 single cell formation Fire Classification BS5852 BS7175 composite Risk assessment to be carried out Risk Factor Very High Risk with pump Electromagnetic Interference...

Страница 14: ...ions along the cells are secure Assess the cells for any punctures which maybe causing a leak Patient is bottoming out Pressure setting may be inadequate adjust comfort range No air is being produced from some of the air outlets of the tube connectors This is normal in alternating mode Air outlets take turns to produce air during their cycle time If fault still continues a Drive DeVilbiss Healthca...

Страница 15: ...nship of our products and our assurance to the customer In the event of a defect covered by this warranty we will at our option supply parts or replace the device This warranty does not cover device failure due to owner misuse or negligence or normal wear and tear The warranty does not extend to non durable components such as rubber accessories castors and grips which are subject to normal wear an...

Страница 16: ...drivedevilbiss co uk The Team at Drive DeVilbiss develops its products to give our customers the freedom to live independently This encompasses their daily home life and provides them with the opportunity to enjoy an outing with family and friends Our goal is to develop a range that will provide individuals with a chance to enjoy everyday life Leaflet code 6002990011 ...

Отзывы: